Operation
4.5.1.4

1 to 3-point calibration

4.5.2

Calibration of the SOLID (TS) parameter

4.5.2.1

Select the SOLID (TS) parameter

22
Note: The turbidity measurement has been calibrated by the manufacturer.
Note: Before the sensor can be calibrated to the TRB parameter, the parameter must be selected
(refer to
4.5.1.1 Select the TURBIDITY (TRB)
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CALIBRATE and confirm.
5. Select CALIBRATE and confirm.
6. Select MEMORY and confirm.
7. Select POINT... (point 1, 2 or 3) and confirm.
Once the calibration point has been recorded by the probe, a mark "<<" is shown after
the point or points that has/have been recorded for approximately 3 seconds.
Note: If the Calibrate menu is closed and then opened again before the calibration is complete,
the "<<" mark is shown again. This indicates that the calibration for this point/these points has
not yet been completed. The old calibration values are still being used.
8. Select the recorded POINT and confirm.
9. Enter the laboratory comparison value and confirm.
To record more calibration points, repeat steps 6 to 9.
10. Go back to the MAIN MENU or the measurement mode display.
The instrument automatically sorts the saved calibration points according to the size of the
calibration values, irrespective of the sequence in which the calibration points were
recorded.
Point 1 is always assigned to the smallest calibration value.
Point 2 is assigned to the next smallest calibration value.
Point 3 is assigned to the largest calibration value.
The value calculated in the laboratory can be corrected at any time by overwriting it.
Before the probe can be calibrated to the SOLID (TS) parameter, the parameter must be
selected.
1. Open the·MAIN MENU.
2. Select SENSOR SETUP and confirm.
3. Select the corresponding sensor and confirm.
4. Select CONFIGURE and confirm.
5. Select PARAMETERS and confirm.
6. Select the parameter TS and confirm.
7. Go back to the MAIN MENU or the measurement mode display.