IMMUNITY
IEC 60601
Compliance
test
test level
Level
Power frequency
30 A/m
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
127
ART38553-101_rev-A_manual.indd 127-128
Electromagnetic
environment – guidance
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
domestic, commercial, or hospital
environment.
IMMUNITY
IEC 60601
Compliance
test
test level
Level
Conducted RF
6 Vrms
6 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Electromagnetic
environment – guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
System, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
1.2
d
=
128