Figure 6‑61 Materials — finding the best frequency using the MIA technique
The following products are used in this procedure:
•
Honeycomb composite reference standard: 25 mm (1 in.) thick, twelve 3-plies
fiberglass top and bottom skins, Nomex and fiberglass cores. Includes two 25 mm
(1 in.) disbonds, two 25 mm (1 in.) delaminations and two 25 mm (1 in.) repairs.
P/N: CHRS-1-3 [U8860626]
•
Cable for pitch-catch and MIA modes, 1.83 m (6 ft).
P/N: SBM-CPM-P11 [U8800058]
•
Right angle MIA probe, 13 mm (0.5 in.) tip. P/N: S-MP-3 [U8010011]
To set the initial BondMaster 600 configuration
1.
Connect the probe and the cable to the PROBE connector of the BondMaster 600.
2.
When prompted, press CONTINUE (A key) to accept the PowerLink probe
information.
NOTE
If you are using a non-PowerLink probe, access the APPL SELECT menu (A key)
using the ADV SETUP menu key (
DMTA-10045-01EN, Rev. E, August 2016
).
Applications
167