ImpOrTANT INFOrmATION rEgArdINg ELECTrO mAgNETIC COmpATIBILITy (EmC)
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Electromagnetic immunity IEC60601-1-2
IEC 60601 Test level
3 V rms
150 kHz to 80 MHz
6 V rms
The ISM bands and the amateur radio
bands between 150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
IEC 60601-1-2:2014 Table 9
(Up to 28 V/m at certain
frequencies)
Compliance level
3 V rms
150 kHz to 80 MHz
6 V rms
The ISM bands and the amateur radio
bands between 150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
IEC 60601-1-2:2014 Table 9
(Up to 28 V/m at certain
frequencies)
47
Electromagnetic environment –guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the PM601
including cables, than the recommended separation
distance calculated from the equation appropriate to
the frequency of the transmitter.
Recommended separation distance
d = 1.16
150 kHz to 80 MHz
d = 0.58
150 kHz to 80 MHz
(The ISM bands and the amateur radio bands)
d = 0.35
80 MHz to 800 MHz
d = 0.7
800 MHz to 2.7 GHz