Omron Avail PM601 Instruction Manual page 47

Wireless dual channel tens
Table of Contents

Advertisement

ImpOrTANT INFOrmATION rEgArdINg ELECTrO mAgNETIC COmpATIBILITy (EmC)
Immunity test
Voltage dips, short
interruptions and voltage
variations on power supply
IEC 61000-4-11
Power frequency (50/ 60
Hz) magnetic field
IEC 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
Electromagnetic immunity IEC60601-1-2
IEC 60601 Test level
0 % U
T
for 0.5 cycle
0 % U
T
for 1 cycle
70 % U
T
for 25 cycles
0 % U
T
for 5 s
30 A/m
Compliance level
0 % U
T
for 0.5 cycle
0 % U
T
for 1 cycle
70 % U
T
for 25 cycles
0 % U
T
for 5 s
30 A/m
46
Electromagnetic environment –guidance
Mains power quality should be that of a typical home,
commercial or hospital environment. If user requires
continued operation during power mains interruption
insure that batteries are installed and charged. Insure
that battery life exceeds longest anticipated power
outrages or provide an additional uninterruptible
power source.
Magnetic fields from common appliances are not
expected to affect the device.