Guidance and manufacturer's declaration – electromagnetic immunity
The ResMed S7 Lightweight is intended for use in the electromagnetic environment specified below. The
customer or the user of the ResMed S7 Lightweight should assure that it is used in such an environment.
IEC60601-1-2
Immunity test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the ResMed S7 Lightweight is
used exceeds the applicable RF compliance level above, the ResMed S7 Lightweight should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the ResMed S7 Lightweight.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications
equipment should be used no closer to any part
of the ResMed S7 Lightweight, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.17 √P
3 Vrms
d = 0.35 √P 80 MHz to 800 MHz
10 V/m
d = 0.70 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
b
YSTEM
PECIFICATIONS
S
S
a
27