Frequency Domain Point-in-Pulse Measurements
This chapter contains measurement calibration and measurement procedures to display the
pulsed-RF S-parameters versus frequency.
Measurement Calibration for Point-in-Pulse
Measurement calibration for point-in-pulse is accomplished in exactly the same way as for the
standard HP 8510.
1. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
4 DOMAIN 5
displayed, select the wideband detector by pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DETECTOR: WIDE BW
2. Select the maximum number of points required for the measurement, then perform the
appropriate measurement calibration.
Note for internal triggering, when you press the key to measure the calibration standard, the
HP 8510 pulse output signal is set to the active state (RF always On) during measurement of
the standard. This assures that the calibration is made with respect to the On portion of the
pulse independent of the trigger delay.
Note for external triggering, you control the pulse width and duty cycle during calibration and
measurement. You must set the trigger delay to make the measurement at the correct time
during the pulse for calibration, then not change it during the measurement.
Measurement
Set Pulse Polarity
After instrument preset, the pulse polarity is set to High for the On period of the pulse
appearing at the HP 8510 rear panel PULSE OUTPUT connector. Use the following
procedure to set the pulse polarity.
Press:
NNNNNNNNNNNNNN
MORE
4 SYSTEM 5
to select the frequency domain. If the W annotation is not
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PULSE CONFIG
PULSE OUT: HIGH
Frequency Domain Point-in-Pulse Measurements 6-1
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MORE
PULSE CONFIG
4 SYSTEM 5
or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PULSE OUT: LOW
6
.