Pulse Profile Domain Measurements
This chapter contains measurement calibration and measurement procedures to display the
pulsed-RF S-parameters versus time.
Measurement Calibration for Pulse Profile
After selection of the pulse prole frequency, measurement calibration for pulse prole
measurements is accomplished in exactly the same way as for the standard HP 8510.
Following are two measurement calibration methods, one for calibration in the Pulse Prole
domain, and the second for calibration using the Frequency List feature. Select the best one
for your application.
The pulse prole domain calibration procedure is an easy way to familiarize yourself with
the pulse prole domain and for general-purpose measurements using simple response-only
correction.
Use the frequency list technique when more than one pulse prole frequency and 1-Port or
2-Port correction is required. Given adequate signal levels, accurate timing, and frequency
stability, there is essentially no dierence in the accuracy of the pulse prole measured data
whether the calibration is performed in the pulse prole domain as described in the rst
procedure or in the frequency domain as described in the second procedure. The main
advantage of calibration in the frequency domain is that you can calibrate at all frequencies in
the list while only connecting the standards once.
Pulse Profile Domain
Perform measurement calibration in the pulse prole domain as follows.
1. Turn correction o by pressing
2. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
4 DOMAIN 5
select the sweep mode.
3. Press
then set the start frequency to the desired pulse prole frequency. This will
4 START 5
be the frequency measured when the pulse prole domain is selected.
4. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PULSE PROFILE
function readout shows both the time value at the marker position and the current
measurement frequency.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION OFF
4 CAL 5
to select the frequency domain. Press
on the Domain menu. Press
.
4 STIMULUS MENU 5
and notice that the active
4 MARKER 5
Pulse Profile Domain Measurements 7-1
7
NNNNNNNNNNNNNN
to
STEP