Maintenance; Maintenance Safety Inspection Checklist - Invacare RPS350–1E User Manual

Stand assist
Hide thumbs Also See for RPS350–1E:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

8 Maintenance

8.1 Maintenance Safety Inspection Checklist

WARNING!
– Maintenance MUST be performed ONLY by
qualified personnel.
– A competent person must ensure that the
strength and stability of the product remains
adequate for all tasks if parts are replaced.
– DO NOT overtighten the mounting hardware,
this will damage the mounting brackets.
Follow the maintenance procedures described in
this manual to keep your patient lift in continuous
service.
The Invacare Patient Lift is designed to provide a maximum
of safe, efficient and satisfactory service with minimum care
and maintenance.
All parts of the Invacare Lift are made of the best grades of
steel, but metal to metal contact will wear after considerable
use.
There is no adjustment or maintenance of the casters, other
than cleaning, lubrication and checking axle and swivel bolts
for tightness. Remove all debris, etc. from the wheel and
swivel bearings. If any parts are worn, replace these parts
IMMEDIATELY.
If you question the safety of any part of the lift, contact
your Dealer IMMEDIATELY.
1078985-I
The UK Health and Safety Executive's Lifting Operations and
Lifting Equipment Regulations 1998 require any equipment
that is used in the workplace to lift a load be subject to
safety inspection on a six monthly basis. This is to be carried
out by a suitably trained individual. Please refer to the HSE
website for guidance (www.hse.gov.uk).
Regular maintenance of patients lifts and accessories is
necessary to assure proper operation. It is the responsibility
of the person responsible for the patient lift to ensure that
this maintenance is carried out.
After the first 12 months of operation, inspect all pivot
points and fasteners for wear. If the metal is worn, the
parts MUST be replaced. Perform this inspection every six
months thereafer.