Table 4 - Guidance and Manufacturer's declaration – Immunity
The HomeSafe/GoSafe are intended for use in the electromagnetic
environment specified below. The customer or user of the HomeSafe/
GoSafe should ensure that it is used in such an environment.
EN/IEC 60601
Immunity Test
Test Level
Conducted RF
3 Vrms
EN/IEC
150 kHz
61000-4-6
to 80 MHz
Radiated RF
10 V/m
EN/IEC
80 MHz
61000-4-3
to 2.5 GHz
* NOTE:
Conducted RF Immunity does not apply to an internal battery operated
devices (7000PHB, 7000AHB, 7000MHB and 7100MHB). Limit above is set for the
battery charger
Compliance Electromagnetic
Environment – Guidance
Level
D=1.17√P
*
3 Vrms
D=0.35√P
80 to 800 MHz
D=0.70√P
800 MHz to 2.5 GHz
Portable and mobile communications
10V/m
equipment should be separated from
the HomeSafe/GoSafe by no less than
the distances calculated/ listed below
D=(3.5/V1)(√P)
D=(3.5/E1)(√P)
80 to 800 MHz
D=(7/Eq)(√P)
800 MHz to 2.5 GHz
where P is the maximum power in
watts (W) and D is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,(a) should
be less than the compliance level in
each frequency range.(b)
Interference may occur in the vicinity
of equipment marked with the
following symbol:
For assistance, call 1-800-635-6156
:
*