In Figures 5-3, 5-6, and 5-9, Fluke recommends using MCS measuring
leads for triaxial shunts and MCP leads for planar shunts. Triaxial shunts
are equipped with guard connectors in the plugs and planar shunts are
equipped with guard sockets.

Measurement with Voltage and Current Transducer

Figure 5-4 shows the connections for measurements with a voltage and current
transducer.
To prevent damage to the transducer due to overload, check
transducer rating.
Transducer errors limit the measuring bandwidth and reduce the intrinsic
uncertainty.
To avoid possible electric shock or personal injury:
Risk of injury when touching connections, internal circuits,
and measuring devices that are not connected to earth
ground.
Always adhere to the instructions regarding the sequence of
connection (see Chapter 5, "Connecting Sequence").
L1
N
Shop for Fluke products online at:
Note
W Caution
Note
XW Warning
P
P
2
1
S
2
S
1
N
n
HI
HI
VOLTAGE
VOLTAGE
1000 V max
1000 V max
LO
LO
3
2
PROBE
PROBE
EXT.SHUNT
EXT.SHUNT
10 V max
10 V max
HI
HI
CURRENT
CURRENT
10 A max
10 A max
LO
LO
ALL INPUTS MAX 1000V CATII TO
Figure 5-4. Voltage and Current Transducer Measurement
MyFlukeStore
www.
A
a
Max. 10 m
V
1
.com
Connection to Circuits
1-Phase Measurement
Load
esn011.eps