Measurement with Star Point Adapter

In systems with three voltage wires, the three wires should be connected to the HI
terminal and the LO terminals should be connected together.
However, in high frequency switching electronics systems such as drives, inverters, and
UPS, this method has additional errors due to high frequency components being shunted
to the ground connection. To compensate for this error, use the star point adapter, see
Figure 5-11.