After installing new batteries, the power up procedure will enable a reset of the battery capacity
calculation and the setup of date and time. Battery capacity reset, date, and time can also be
corrected via functions FT508 and FT200.
When the battery plug is connected, the meter will display the meter version for 10 seconds. The
display will then show "rESEt.bAt" indicating the option to reset the internal battery power
calculation. To execute the reset, press the key within 6 seconds. If the key is not pressed, the
meter will proceed to the set date and set clock and finally normal operation mode.
If the key is pressed within the reset battery time, the display will indicate "Accept" to ensure that
the reset should take place. The reset function will take place only if the key is pressed again within
the next 6 seconds. If not, normal operation will begin.
For setting up the date and time, the different key function must be used - see section 5.2 „Operator
menu" index 1. An "A" indicates an acceptable value and a flashing "A" indicates the value is
stored when the key if released.
The reset function also sets the actual date as the battery replacement date.