MAG 8000 & MAG 8000 CT
Connection diagram for
115/230 V AC (mains) or
12-24 V AC/DC (line)
power supply
115/230 V AC (mains)
power supply
12-24 V AC/DC (line)
power supply
12
3. Electrical connection
Mains power input:
Mains power input:
Mains power output:
Battery backup input:
Functional ground:
Mains power supply has to be connected to a switch, near the flowmeter, according to IEC 61010-1
clause 5.4.3.d
Line power input:
Line power input:
Line power output:
Battery backup input:
Functional ground:
SFIDK.PS.026.D9.02
Factory mounted PUR cable with 2 x 1 mm
cable length = 3 m
Brown wire - L (line, hot) and blue wire - N (neutral, cold)
Battery connector - female with blue wire and yellow wire, blue wire is
ground. The battery connector - female has to be connected to the
male connector 3.6 V DC on the PCB board
Battery connector - male with black wire and red wire, black wire is ground.
The battery connector - male has to be connected to the female
connector on the backup battery
Black wire with terminal must be connected to MAG 8000 or MAG 8000 CT
encapsulation with a screw
Factory mounted PUR cable with 2 x 1 mm
cable length = 3 m
Brown wire - L (line, hot, positive) and blue wire (neutral, cold, negative)
Battery connector - female with blue wire and yellow wire, blue wire is
ground. The battery connector - female has to be connected to the
male connector 3.6 V DC on the PCB board
Battery connector - male with black wire and red wire, black wire is ground.
The battery connector - male has to be connected to the female
connector on the backup battery
Black wire with terminal must be connected to MAG 8000 or MAG 8000 CT
encapsulation with a screw