Siemens SITRANS F M MAG 8000 Operating Manual page 9

Battery operated water meter
Hide thumbs Also See for SITRANS F M MAG 8000:
Table of Contents

Advertisement

MAG 8000 & MAG 8000 CT
Maximum allowable
torques
(continued)
Remote installation
Installation of add-on
module
2. Installation
Nominal size
mm
inch
25
1"
40
1½"
50
2"
65
2½"
80
3"
100
4"
125
5"
150
6"
200
8"
250
10"
300
12"
350
14"
400
16"
450
18"
500
20"
600
24"
Verify that the model and serial numbers shown on the labels of the sensor and transmitter are
matched properly (1). Make sure that the cable is safety installed to avoid damaging of cable and
connectors. Please note the different connector types for the coil and electrodes, both having a
minimum diameter of 90 mm (3.6 inches). Save the dust covers for future use and protection (2).
Make sure the connectors are clean and fastened securely to achieve a good connection and
watertight seal (3 & 4).
Note
If dirt enters the connector ends, use plain water for cleaning. Ensure the connectors are
completely dry before making connections.
SFIDK.PS.026.D9.02
PN 10
PN 16
Nm
f/lbs
Nm
f/lbs
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
25
18
N/A
N/A
25
18
N/A
N/A
25
18
N/A
N/A
25
18
N/A
N/A
29
21
N/A
N/A
50
37
50
37
50
37
50
37
82
61
57
42
111
82
60
44
120
89
88
65
170
125
92
68
170
125
103
76
230
170
161
119
350
258
The module must be mounted on the backside
of the MAG 8000 or MAG 8000 CT electronic.
Use the two 3 mm screws and washers to fix the
module to the MAG 8000 or MAG 8000 CT
electronic, with maximum torque 0.5 Nm.
PN 40
Class 150
Nm
f/lbs
Nm
f/lbs
10
7
7
5
16
12
9
7
N/A
N/A
25
18
N/A
N/A
25
18
N/A
N/A
34
25
N/A
N/A
26
19
N/A
N/A
42
31
N/A
N/A
57
42
N/A
N/A
88
65
N/A
N/A
99
73
N/A
N/A
132
97
N/A
N/A
225
166
N/A
N/A
210
155
N/A
N/A
220
162
N/A
N/A
200
148
N/A
N/A
280
207
9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans f m mag 8000 ct

Table of Contents