Siemens SITRANS F M MAG 8000 Operating Manual page 6

Battery operated water meter
Hide thumbs Also See for SITRANS F M MAG 8000:
Table of Contents

Advertisement

MAG 8000 & MAG 8000 CT
Installation in large pipes
(MAG 8000)
Installation instructions
6
2. Installation
SFIDK.PS.026.D9.02
The water meter can be installed between two
reducers (e.g. DIN 28545). With an 8° reducer,
the following pressure drop curve applies. The
curves are applicable to water.
Example:
A flow velocity of 3 m/s (10 ft./sec.) (V) in a
sensor with a diameter reduction from DN 100
to DN 80 (4" to 3") (d
/d
= 0.8) gives a pressure
1
2
drop of 2.9 mbar (0.04 psi).
The sensor must always be completely full with
liquid.
Therefore avoid:
• Air in the pipe
• Installation at the highest point in the pipe
system
• Installation in vertical pipes with free outlet.
For partially filled pipes or pipes with downward
flow and free outlet, the flowmeter should be
located in a U-tube.