3. PREPARATION AND SAFETY CHECK

3-1 PREPARATION

WARNING - PENTAX video processors are electro-medical devices incorporating delicate
components and sophisticated circuitry that should not be subjected to harsh conditions,
including excessive vibrations and/or severe impact. NEVER drop this equipment or subject it to
severe impact as it could compromise the functionality and /or safety of the unit. Should this
equipment be mishandled or dropped, do NOT use it. Return it to an authorized PENTAX service
facility for inspection and repair.

(1) SETTING UP THE VIDEO PROCESSOR

1. Place the video processor on a stable and level surface (cart, counter, stand,etc..).
NOTE
• Avoid places where the video processor may be splashed with liquid.
• Absolutely do NOT use in any environment with explosive or flammable gases.
• Avoid places where the unit could be exposed to high temperature, humidity, direct sun light, etc.
• Do NOT install, operate or store electro-medical equipment in a dusty environment.
Accumulation of dust within these units may cause malfunction, smoke, or ignition.
• Do NOT block the venting grids on the sides of the processor.
• When moving the video processor, do NOT hold the scope locking lever.
2. Make sure the power switch is OFF.
3. Plug the power cord into the power source
using the three (3) prong plug supplied with the unit.
4. Ensure the keyboard is connected properly.
WARNING
• To reduce the potential for the electric shock, connect the power cord
of the following peripheral components into the medical isolation
transformer supplied.
– Any peripheral components connected to the PENTAX Processor.
– Any peripheral components connected to the PENTAX Endoscope.
– NEVER connect other components to the isolation transformer.
• Check that the total power consumption of all connected devices does
not exceed the isolation transformers' power rating.
• Make sure that outputs are in compliance with IEC 60601-1-1.
• Make sure that the power cord is connected to the main with a three (3)
prong plug (In the U.S.A use UL 2601-1 rated isolation transformers
and/or power strips only).

(2) CONNECTING THE WATER BOTTLE

1. Fill the water bottle approximately 2/3 full with sterile water.
2. Screw the water bottle cap assembly to the water bottle snugly.
-Do NOT overtighten the water bottle cap
NOTE
3. Set the Air/Water- Drain lever to A/W position.
4. Insert the water bottle air pipe stem into the video processor water
bottle receptacle and press until the water bottle 'clicks' into
position.
-Do NOT press the water bottle too forcefully into the video
NOTE
processor. Rough handling may cause water to leak onto/into the
video processor.
5. Insert the Air/Water connector into the holder on the water bottle
cap assembly until the endoscope is connected
NOTE
-Always disconnect the water bottle before moving the
processor into a position not common to normal use.
Always disconnect the water bottle before packing the video
processor for shipment.