Device functions Proline Prosonic Flow 93 1 Notes on using this manual Notes on using this manual There are various ways of locating the description of a function of your choice in the manual: Using the table of contents to locate a function description The designations of all the cells in the function matrix are listed in the table of contents.
2 Function matrix Device functions Proline Prosonic Flow 93 Function matrix General layout of the function matrix The function matrix consists of four levels: Blocks -> Groups -> Function groups -> Functions 0000 0001 0002 0003 0009 … 0400 0401...
Device functions Proline Prosonic Flow 93 2 Function matrix 2.1.5 Codes identifying cells Each cell (block, group, function group and function) in the function matrix has an individual, unique code. Blocks: The code is a letter (A, B, C, etc.) Groups: The code consists of three letters (AAA, ABA, BAA, etc.).
2 Function matrix Device functions Proline Prosonic Flow 93 Function matrix Proline Prosonic Flow 93 FUNCTION BLOCKS GROUPS GROUPS → → MEASURED VARIABLES MEASURING VALUES see P. 12 → (see P. 11) SYSTEM UNITS see P. 15 ↓ → SPECIAL UNITS see P.
3 Block MEASURED VARIABLES Device functions Proline Prosonic Flow 93 Group MEASURING VALUES 3.1.1 Function group MAIN VALUES CH1 ⇒ ⇒ MEASURED VARIABLES MEASURING VALUES MAIN VALUES CH1 000 Functional description ➞ ➞ MEASURED VARIABLES MEASURING VALUES MAIN VALUES CH1 The measuring values of channel 1 currently being measured are displayed in this function group.
Device functions Proline Prosonic Flow 93 3 Block MEASURED VARIABLES 3.1.2 Function group MAIN VALUES CH2 ⇒ ⇒ MEASURED VARIABLES MEASURING VALUES MAIN VALUES CH2 Functional description ➞ ➞ MEASURED VARIABLES MEASURING VALUES MAIN VALUES CH2 The measuring values of channel 2 currently being measured are displayed in this function group.
3 Block MEASURED VARIABLES Device functions Proline Prosonic Flow 93 3.1.3 Function group CALCULATED MAIN VALUES ⇒ ⇒ MEASURED VARIABLES MEASURING VALUES CALCUL. MAIN VALUES 008 Functional description ➞ ➞ MEASURED VARIABLES MEASURING VALUES CALCULATED MAIN VALUES The calculated measured values appear on the display. The measured values of both channels are used when calculating the values.
Device functions Proline Prosonic Flow 93 3 Block MEASURED VARIABLES Group SYSTEM UNITS 3.2.1 Function group CONFIGURATION ⇒ MEASURED VARIABLES MEASURING VALUES ⇓ ⇒ SYSTEM UNITS CONFIGURATION Functional description ➞ ➞ MEASURED VARIABLES SYSTEM UNITS CONFIGURATION You can select the units for measured variables in this function group.
Page 16
3 Block MEASURED VARIABLES Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ MEASURED VARIABLES SYSTEM UNITS CONFIGURATION UNIT VOLUME Use this function to select the unit for displaying the volume. (0403) The unit you select here is also valid for: •...
Device functions Proline Prosonic Flow 93 3 Block MEASURED VARIABLES 3.2.2 Function group ADDITIONAL CONFIGURATION ⇒ MEASURED VARIABLES MEASURING VALUES ⇓ ⇒ SYSTEM UNITS CONFIGURATION ⇓ ADDIT. CONFIG. Functional description ➞ ➞ MEASURED VARIABLES SYSTEM UNITS ADDITIONAL CONFIGURATION UNIT TEMPERATURE Use this function to select the unit for displaying the liquid temperature.
3 Block MEASURED VARIABLES Device functions Proline Prosonic Flow 93 Group SPECIAL UNITS 3.3.1 Function group ARBITRARY UNIT ⇒ MEASURED VARIABLES MEASURING VALUES ⇓ SYSTEM UNITS ⇓ ⇒ SPECIAL UNITS ARBITRARY UNIT Functional description ➞ ➞ MEASURED VARIABLES SPECIAL UNITS ARBITRARY UNIT Use this function group to define an arbitrary unit for the flow rate variable.
Factory setting: Note! You will find a flowchart of the Quick Setup SENSOR menu on Page 21. Please refer to the Operating Instructions Proline Prosonic Flow 93 (BA 070D/06/en/..) for more information on Quick Setup menus. QUICK SETUP Use this function to start the Quick Setup menu for commissioning.
Page 20
4 Block QUICK SETUP Device functions Proline Prosonic Flow 93 Functional description QUICK SETUP T-DAT SAVE/LOAD Use this function to save the parameter settings / configuration of the transmitter in a (1009) transmitter DAT (T-DAT), or to load the parameter settings from the T-DAT into the EEPROM (manual security function).
Device functions Proline Prosonic Flow 93 4 Block QUICK SETUP Quick Setup “Sensor” If the measuring device is equipped with a local operation, the sensor distance can be ascertained using the “Sensor” Quick Setup menu. If a measuring device does not have a local operation, the individual parameters and functions must be configured via the configuration program ToF Tool - Fieldtool Package.
Page 22
4 Block QUICK SETUP Device functions Proline Prosonic Flow 93 Note! • The display returns to the cell SETUP SENSOR (1001) if you press the ESC key combination (X) during interrogation. • If you confirm the inquiry “Automatic configuration of the display?” as YES, the configuration of the display will be carried out as follows: main line = volume flow, additional line = totalizer 1, information line = operational/system condition.
Device functions Proline Prosonic Flow 93 4 Block QUICK SETUP Quick Setup “Commissioning” If the measuring device is equipped with a local operation, all the device parameters important for standard measuring mode can be configured easily and quickly using the “Commissioning” Quick Setup menu.
Page 24
4 Block QUICK SETUP Device functions Proline Prosonic Flow 93 Note! • The display returns to the cell QUICK SETUP COMMISSIONING (1002) if you press the ESC key combination (X) during interrogation. • If you confirm the inquiry “Automatic configuration of the display?” as YES, the configuration of the display will be carried out as follows: main line = volume flow, additional line = totalizer 1, information line = operational/system condition.
Device functions Proline Prosonic Flow 93 4 Block QUICK SETUP Pulsating flow Quick Setup menu If the measuring device is equipped with a local operation, all the device parameters important for standard measuring mode can be configured easily and quickly using the “Pulsating Flow” Quick Setup menu.
Page 26
4 Block QUICK SETUP Device functions Proline Prosonic Flow 93 Note! • The display returns to the cell QUICK SETUP PULSATING FLOW (1003) if you press the ESC key combination (X) during interrogation. • You can call this Quick Setup menu either directly from the “COMMISSIONING” Quick Setup menu or manually by means of the QUICK SETUP PULSATING FLOW function (1003).
“ENGLISH”. • You can change the language group via the configuration software ToF Tool - Fieldtool Package. Please do not hesitate to contact your Endress+Hauser sales office if you have any questions. DISPLAY DAMPING...
Page 29
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE Functional description ➞ ➞ USER INTERFACE CONTROL BASIC CONFIGURATION CONTRAST LCD Use this function to optimize display contrast to suit local operating conditions. (2003) User input: 10...100% Factory setting: BACKLIGHT Use this function to optimize the backlight to suit local operating conditions.
HOME position. • You can also disable programming in this function by entering any number (other than the defined private code). • The Endress+Hauser service organization can be of assistance if you mislay your personal code. DEFINE PRIVATE CODE Use this function to enter a personal code number for enabling programming.
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE 5.1.3 Function group OPERATION ⇒ ⇒ USER INTERFACE CONTROL BASIC CONFIGURATION 200 ⇓ UNLOCKING/LOCKING 202 ⇓ OPERATION Functional description ➞ ➞ USER INTERFACE CONTROL OPERATION TEST DISPLAY Use this function to test the operability of the local display and its pixels.
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 Group MAIN LINE 5.2.1 Function group CONFIGURATION ⇒ USER INTERFACE CONTROL ⇓ ⇒ MAIN LINE CONFIGURATION Functional description ➞ ➞ USER INTERFACE MAIN LINE CONFIGURATION A0001253 1 = main line, 2 = additional line, 3 = information line...
Page 33
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE Functional description ➞ ➞ USER INTERFACE MAIN LINE CONFIGURATION 100% VALUE Note! (2201) This function is only available if VOLUME FLOW IN % or CALCULATED VOLUME FLOW IN % is selected in the function ASSIGN (2200).
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 5.2.2 Function group MULTIPLEX ⇒ USER INTERFACE CONTROL ⇓ ⇒ MAIN LINE CONFIGURATION ⇓ MULTIPLEX Functional description ➞ ➞ USER INTERFACE MAIN LINE MULTIPLEX ASSIGN Use this function to define a second reading to be displayed in the main line alternatively (2220) (every 10 seconds) with the reading defined in the function ASSIGN (2200).
Page 35
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE Functional description ➞ ➞ USER INTERFACE MAIN LINE MULTIPLEX FORMAT Use this function to define the maximum number of places after the decimal point for the (2222) second value displayed in the main line.
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 Group ADDITIONAL LINE 5.3.1 Function group CONFIGURATION ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ⇒ ADDITIONAL LINE CONFIGURATION Functional description ➞ ➞ USER INTERFACE ADDITIONAL LINE CONFIGURATION A0001253 1 = main line, 2 = additional line, 3 = information line...
Page 37
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE Functional description ➞ ➞ USER INTERFACE ADDITIONAL LINE CONFIGURATION ASSIGN Advanced options with optional software package ADVANCED DIAGNOSTICS: (continued) DEVIATION VOLUME FLOW (CH1...CH2) DEVIATION VOLUME FLOW AVERAGE DEVIATION FLOW VELOCITY (CH1...CH2) DEVIATION FLOW VELOCITY AVERAGE DEVIATION SIGNAL STRENGTH (CH1...CH2)
Page 38
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ USER INTERFACE ADDITIONAL LINE CONFIGURATION DISPLAY MODE Note! (2403) This function is only available if VOLUME FLOW BARGRAPH IN % or CALCULATED VOLUME FLOW BARGRAPH IN % was selected in the function ASSIGN (2420).
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE 5.3.2 Function group MULTIPLEX ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ⇒ ADDITIONAL LINE CONFIGURATION ⇓ MULTIPLEX Functional description ➞ ➞ USER INTERFACE ADDITIONAL LINE MULTIPLEX ASSIGN Use this function to define the second reading to be displayed in the additional line (2420) alternately (every 10 seconds) with the value defined in the function ASSIGN (2400).
Page 40
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ USER INTERFACE ADDITIONAL LINE MULTIPLEX ASSIGN Advanced options with optional software package ADVANCED DIAGNOSTICS: (continued) DEVIATION VOLUME FLOW (CH1...CH2) DEVIATION VOLUME FLOW AVERAGE DEVIATION FLOW VELOCITY (CH1...CH2) DEVIATION FLOW VELOCITY AVERAGE DEVIATION SIGNAL STRENGTH (CH1...CH2)
Page 41
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE Functional description ➞ ➞ USER INTERFACE ADDITIONAL LINE MULTIPLEX DISPLAY MODE Note! (2423) This function is only available if VOLUME FLOW BARGRAPH IN % or CALCULATED VOLUME FLOW BARGRAPH IN % was selected in the function ASSIGN (2420).
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 Group INFORMATION LINE 5.4.1 Function group CONFIGURATION ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ADDITIONAL LINE ⇓ ⇒ INFORMATION LINE CONFIGURATION Functional description ➞ ➞ USER INTERFACE INFORMATION LINE...
Page 43
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE Functional description ➞ ➞ USER INTERFACE INFORMATION LINE CONFIGURATION ASSIGN Advanced options with optional software package ADVANCED DIAGNOSTICS: (continued) DEVIATION VOLUME FLOW (CH1...CH2) DEVIATION VOLUME FLOW AVERAGE DEVIATION FLOW VELOCITY (CH1...CH2) DEVIATION FLOW VELOCITY AVERAGE DEVIATION SIGNAL STRENGTH (CH1...CH2)
Page 44
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ USER INTERFACE INFORMATION LINE CONFIGURATION DISPLAY MODE Note! (2603) This function is only available if VOLUME FLOW BARGRAPH IN % or CALCULATED VOLUME FLOW BARGRAPH IN % was selected in the function ASSIGN (2600).
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE 5.4.2 Function group MULTIPLEX ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ADDITIONAL LINE ⇓ ⇒ INFORMATION LINE CONFIGURATION ⇓ MULTIPLEX Functional description ➞ ➞ USER INTERFACE INFORMATION LINE MULTIPLEX...
Page 46
5 Block USER INTERFACE Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ USER INTERFACE INFORMATION LINE MULTIPLEX ASSIGN Advanced options with optional software package ADVANCED DIAGNOSTICS: (continued) DEVIATION VOLUME FLOW (CH1...CH2) DEVIATION VOLUME FLOW AVERAGE DEVIATION FLOW VELOCITY (CH1...CH2) DEVIATION FLOW VELOCITY AVERAGE DEVIATION SIGNAL STRENGTH (CH1...CH2)
Page 47
Device functions Proline Prosonic Flow 93 5 Block USER INTERFACE Functional description ➞ ➞ USER INTERFACE INFORMATION LINE MULTIPLEX DISPLAY MODE Note! (2623) This function is only available if VOLUME FLOW BARGRAPH IN % or CALCULATED VOLUME FLOW BARGRAPH IN % was selected in the function ASSIGN (2620).
Device functions Proline Prosonic Flow 93 6 Block TOTALIZER Group TOTALIZER (1...3) 6.1.1 Function group CONFIGURATION ⇒ ⇒ TOTALIZER TOTALIZER 1 CONFIGURATION ⇓ ⇒ TOTALIZER 2 CONFIGURATION ⇓ ⇒ TOTALIZER 3 CONFIGURATION Functional description ➞ ➞ TOTALIZER TOTALIZER (1...3) CONFIGURATION The function descriptions below apply to totalizers 1...3;...
Page 50
6 Block TOTALIZER Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ TOTALIZER TOTALIZER (1...3) CONFIGURATION TOTALIZER MODE Use this function to define how the flow components are to be totaled by the totalizer in (3002) question. Options: BALANCE Positive and negative flow components.
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Group CURRENT OUTPUT (1...3) 7.1.1 Function group CONFIGURATION ⇒ ⇒ OUTPUTS CURRENT OUTPUT 1 CONFIGURATION ⇓ ⇒ CURRENT OUTPUT 2 CONFIGURATION ⇓ ⇒ CURRENT OUTPUT 3 CONFIGURATION Functional description ➞ ➞...
Page 55
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION CURRENT SPAN Use this function to define the current span. The selection specifies the operational range (4001) and the lower and upper signal on alarm. For the current output 1 the option HART can be defined additionally.
Page 56
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION VALUE 0_4 mA Use this function to assign the 0/4 mA current a value. The value can be greater or less (4002) than the value assigned to 20 mA (function VALUE 20 mA (4003)).
Page 57
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION VALUE 0_4 mA Parameter setting example A: (Continued) 1. VALUE 0_4 mA (4002) = not equal to zero flow (e.g. -5 m VALUE 20 mA (4003) = not equal to zero flow (e.g.
Page 58
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION VALUE 20 mA Use this function to assign the 20 mA current a value. The value can be greater or less (4003) than the value assigned to 0/4 mA, (function VALUE 0_4 mA (4002)).
Page 59
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION MEASURING MODE Use this function to define the measuring mode for the current output. (4004) Options: STANDARD SYMMETRY PULSATING FLOW The current output signal is proportional to the measured variable. The flow components ➀...
Page 60
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION MEASURING MODE PULSATING FLOW (Continued) If flow is characterized by severe fluctuations as is the case, for example, with recipro- cating pumps, flow components outside the measuring range are buffered, balanced and output after a maximum delay of 60 seconds.
Page 61
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION ➀ ➁ ➀ ➁ Detailed explanations and 2. Defined measuring range ( – do not have the same sign. information (Continued) ➀...
Page 62
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS CURRENT OUTPUT (1...3) CONFIGURATION FAILSAFE MODE For safety reasons it is advisable to ensure that the current output assumes a predefined (4006) state in the event of a fault. The setting you select here affects only the current output. It has no effect on other outputs and the display (e.g.
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Group PULSE/FREQUENCY OUTPUT (1...2) 7.2.1 Function group CONFIGURATION ⇒ OUTPUTS CURRENT OUTPUT 1 ⇓ CURRENT OUTPUT 2 ⇓ CURRENT OUTPUT 3 ⇓ ⇒ PULSE/FREQ. OUTP. 1 CONFIGURATION ⇓ ⇒ PULSE/FREQ. OUTP. 2...
Page 66
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) START VALUE Note! FREQUENCY This function is not available unless the FREQUENCY setting was selected in the (4202) OPERATION MODE function (4200).
Page 67
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) VALUE f LOW Note! (4204) This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 68
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) VALUE f HIGH Parameter setting example A: (Continued) 1. VALUE f LOW (4204) = not equal to zero flow (e.g. -5 m VALUE f HIGH (4205) = not equal to zero flow (e.g.
Page 69
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) MEASURING MODE Note! (4206) This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 70
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) MEASURING MODE • PULSATING FLOW (Continued) If flow is characterized by severe fluctuations as is the case, for example, with recipro- cating pumps, flow components outside the measuring range are buffered, balanced and output after a maximum delay of 60 seconds.
Page 71
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) OUTPUT SIGNAL Note! (4207) Function is not available unless the FREQUENCY setting was selected in the OPERATING MODE (4200) function. For selecting the output configuration of the frequency output.
Page 72
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) OUTPUT SIGNAL Example for output configuration PASSIVE-POSITIVE: (Continued) Output configuration with an external pull-down resistance. In the quiescent state (at zero flow), a positive voltage level is measured via the pull-down resistance.
Page 73
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) OUTPUT SIGNAL Example for active output circuit (ACTIVE): (Continued) With an active circuit, the internal power supply is 24 V. The frequency output is short-circuit proof.
Page 74
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (GENERAL/FREQUENCY) TIME CONSTANT Note! (4208) This function is not available unless the FREQUENCY setting was selected in the OPERATION MODE function (4200).
Page 75
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE) ASSIGN PULSE Note! (4221) This function is not available unless the PULSE setting was selected in the function OPERATION MODE (4200).
Page 76
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE) PULSE WIDTH Note! (4223) This function is not available unless the PULSE setting was selected in the function OPERATION MODE (4200).
Page 77
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE) MEASURING MODE Note! (4225) This function is not available unless the PULSE setting was selected in the function OPERATION MODE (4200).
Page 78
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE) OUTPUT SIGNAL Note! (4226) Function is not available unless the PULSE setting was selected in the OPERATION MODE (4200) function. For selecting the output configuration of the pulse output.
Page 79
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE) OUTPUT SIGNAL Example for output configuration PASSIVE-POSITIVE: (Continued) Output configuration with an external pull-down resistance. In the quiescent state (at zero flow), a positive voltage level is measured via the pull-down resistance.
Page 80
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE) OUTPUT SIGNAL Example for active output circuit (ACTIVE): (Continued) With an active circuit, the internal power supply is 24 V. The pulse output is short-circuit proof.
Page 81
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE) FAILSAFE MODE Note! (4227) This function is not available unless the PULSE setting was selected in the function OPERATION MODE (4200).
Page 82
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (STATUS) ASSIGN STATUS Note! (4241) This function is not available unless the STATUS setting was selected in the function OPERATION MODE (4200).
Page 83
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (STATUS) ON-VALUE Note! (4242) This function is not available unless STATUS was selected in the function OPERATION MODE (4200) and LIMIT VALUE or FLOW DIRECTION was selected in the function ASSIGN STATUS (4241).
Page 84
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (STATUS) OFF-VALUE Note! (4244) This function is not available unless STATUS was selected in the function OPERATION MODE (4200) and a LIMIT VALUE was selected in the function ASSIGN STATUS (4241).
Page 85
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (STATUS) MEASURING MODE Note! (4246) • This function is not available unless the STATUS setting was selected in the function OPERATION MODE (4200).
Page 86
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (STATUS) TIME CONSTANT Note! (4247) This function is not available unless the STATUS setting was selected in the function OPERATION MODE (4200).
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS 7.2.2 Function group OPERATION ⇒ OUTPUTS CURRENT OUTPUT 1 ⇓ CURRENT OUTPUT 2 ⇓ CURRENT OUTPUT 3 ⇓ ⇒ PUL./FREQ. OUTPUT 1 CONFIGURATION ⇓ ⇓ OPERATION ⇒ PUL./FREQ. OUTPUT 2 CONFIGURATION ⇓...
Page 88
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (FREQUENCY) VALUE SIMULATION Note! FREQUENCY This function is not available unless FREQUENCY was selected in the function (4303) OPERATION MODE (4200) and the function SIMULATION FREQUENCY (4302) is active (= ON).
Page 89
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (PULSE) SIMULATION PULSE Note! (4322) This function is not available unless the PULSE option was selected in the OPERATING MODE function. Use this function to activate simulation of the pulse output.
Page 90
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (STATUS) ACTUAL STATUS Note! (4341) This function is not available unless the STATUS setting was selected in the function OPERATION MODE (4200).
• The relay output is configured as a normally open (NO or make) contact by default. It can be reconfigured as a normally closed (NC or break) contact by means of a jumper on the relay module, (Operating Instructions Proline Prosonic Flow 93, BA 070D/06/en/.).
Page 94
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ OUTPUTS RELAY OUTPUT (1...2) CONFIGURATION SWITCH-ON DELAY Note! (4702) This function is not available unless LIMIT VALUE or FLOW DIRECTION was selected in the function ASSIGN RELAY (4700).
Page 95
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS RELAY OUTPUT (1...2) CONFIGURATION MEASURING MODE Note! (4705) This function is not visible unless a limit value was assigned to the relay output. Use this function to define the measuring mode for the relay output.
Use this function to check the current status of the relay output. (4740) A jumper on the contact side defines the relay output as a normally open (NO or make) or normally closed (NC or break) contact Operating Instructions Proline Prosonic Flow 93, BA 070D/06/en/. User interface:...
Page 97
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Functional description ➞ ➞ OUTPUTS RELAY OUTPUT (1...2) OPERATION VALUE SIMULATION Note! SWITCH POINT The function is not visible unless the function SIMULATION SWITCH POINT (4741) is (4742) active (= ON).
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS 7.3.4 Information on the response of the relay output General If you have configured the relay output signal for “LIMIT-VALUE” or “FLOW DIRECTION”, you can define the requisite switching points in the ON-VALUE and OFF-VALUE functions. When the measured variable in question reaches one of these predefined values, the relay output switches as shown in the illustrations below.
7 Block OUTPUTS Device functions Proline Prosonic Flow 93 7.3.5 Switching action of the relay output Contact* Function Status Relais coil ON (operation) System in measuring mode energized A0001239 A0001237 System not in measuring mode (power supply failed) de-energized A0001240...
Page 101
Device functions Proline Prosonic Flow 93 7 Block OUTPUTS Contact* Function Status Relais coil Limit value Limit value not overshot or – Volume flow undershot – Totalizer energized – Sound velocity – Flow velocity A0001243 A0001239 A0001237 (CH1, CH2 Limit value overshot or AVG.
Device functions Proline Prosonic Flow 93 8 Block INPUTS Group STATUS INPUT 8.1.1 Function group CONFIGURATION ⇒ ⇒ INPUTS STATUS INPUT CONFIGURATION Functional description ➞ ➞ INPUTS STATUS INPUT CONFIGURATION ASSIGN STATUS INPUT Use this function to assign a switching function to the status input.
8 Block INPUTS Device functions Proline Prosonic Flow 93 8.1.2 Function group OPERATION ⇒ ⇒ INPUTS STATUS INPUT CONFIGURATION ⇓ OPERATION Functional description ➞ ➞ INPUTS STATUS INPUT OPERATION ACTUAL Use this function to view the current level of the status input.
Device functions Proline Prosonic Flow 93 8 Block INPUTS 8.1.3 Function group INFORMATION ⇒ ⇒ INPUTS STATUS INPUT CONFIGURATION ⇓ OPERATION ⇓ INFORMATION Functional description ➞ ➞ INPUTS STATUS INPUT INFORMATION TERMINAL NUMBER Use this function to display the numbers of the terminals (in the connection compart- (5080) ment) which are used by the status input.
(6004) User interface: OFF (Data exchange is possible) ON (Data exchange is disabled) Factory setting: Note! Write protection is activated and deactivated by means of a jumper on the I/O module Operating Instructions Proline Prosonic Flow 93, BA 070D/06/en/. Endress+Hauser...
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 9.1.2 Function group INFORMATION ⇒ ⇒ BASIC FUNCTION HART CONFIGURATION ⇓ OPERATION ⇓ INFORMATION Functional description ➞ ➞ BASIC FUNCTION HART OPERATION MANUFACTURER ID Use this function to view the manufacturer number in decimal numerical format.
Device functions Proline Prosonic Flow 93 9 Block BASIC FUNCTION Group PROCESS PARAMETER (CH1...CH2) 9.2.1 Function group CONFIGURATION ⇒ BASIC FUNCTION HART ⇓ ⇒ PROCESS PARAM. CH1 CONFIGURATION ⇓ ⇒ PROCESS PARAM. CH2 CONFIGURATION Functional description ➞ ➞ BASIC FUNCTION PROCESS PARAMETER (CH1...CH2)
Page 110
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ BASIC FUNCTION PROCESS PARAMETER (CH1...CH2) CONFIGURATION PRESSURE SHOCK SUP- The closure of a valve can cause brief but severe movements of the fluid in the piping sys- PRESSION tem, movements which the measuring system registers.
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 9.2.3 Function group PIPE DATA ⇒ BASIC FUNCTION HART ⇓ ⇒ PROCESS PARAM. CH1 CONFIGURATION ⇓ ADJUSTMENT ⇓ ⇓ PIPE DATA ⇒ PROCESS PARAM. CH2 CONFIGURATION ⇓ ADJUSTMENT ⇓ PIPE DATA Functional description ➞...
Page 113
Note! • To measure the sound velocity, you require the ultrasonic sensors “DDU18” which you can order as an accessory from Endress+Hauser. • A reference value is used as a basis for measuring the sound velocity. This reference value can be edited. (see function REFERENCE VALUE (6523), page 113) User input: Fixed-point number 800...6500 m/s...
Page 114
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ BASIC FUNCTION PROCESS PARAMETER (CH1...CH2) PIPE DATA CIRCUMFERENCE This function displays the outer circumference of the pipe determined via the value (6525) entered in the function NOMINAL DIAMETER (6521). If you edit the predetermined value the pipe standard will be reset to the option OTHERS and the function NOMINAL DIAMETER (6521) does not appear.
Page 115
Device functions Proline Prosonic Flow 93 9 Block BASIC FUNCTION Functional description ➞ ➞ BASIC FUNCTION PROCESS PARAMETER (CH1...CH2) PIPE DATA LINER MATERIAL This function displays the liner material of the pipe determined via the value entered in (6528) the function PIPE STANDARD (6520). the pipe standard will be reset to the option OTHERS and the function NOMINAL DIAMETER (6521) does not appear.
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 9.2.4 Function group LIQUID DATA ⇒ BASIC FUNCTION HART ⇓ ⇒ PROCESS PARAM. CH1 CONFIGURATION ⇓ ADJUSTMENT ⇓ ⇓ PIPE DATA ⇓ LIQUID DATA ⇒ PROCESS PARAM. CH2 CONFIGURATION ⇓...
Page 117
Device functions Proline Prosonic Flow 93 9 Block BASIC FUNCTION Functional description ➞ ➞ BASIC FUNCTION PROCESS PARAMETER (CH1...CH2) LIQUID DATA SOUND VELOCITY This function displays the sound velocity of the liquid determined via the values entered LIQUID in the functions LIQUID (6540) and TEMPERATURE (6541). If you edit the predeter- (6542) mined value the function LIQUID (6540) will be reset to the option OTHERS.
Page 118
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ BASIC FUNCTION PROCESS PARAMETER (CH1...CH2) LIQUID DATA SOUND VELOCITY Use this function to specify the lower search range for the sound velocity of the liquid. NEGATIVE...
Device functions Proline Prosonic Flow 93 9 Block BASIC FUNCTION Group SYSTEM PARAMETER (CH1...CH2) 9.3.1 Function group CONFIGURATION ⇒ BASIC FUNCTION HART ⇓ PROCESS PARAM. CH1 ⇓ PROCESS PARAM. CH2 ⇓ ⇒ SYSTEM PARAM. CH1 CONFIGURATION ⇓ ⇒ SYSTEM PARAM. CH2...
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 Group SENSOR DATA (CH1...CH2) 9.4.1 Function group SENSOR PARAMETER ⇒ BASIC FUNCTION HART ⇓ PROCESS PARAM. CH1 ⇓ PROCESS PARAM. CH2 ⇓ SYSTEM PARAM. CH1 ⇓ SYSTEM PARAM. CH2 ⇓...
Page 121
Device functions Proline Prosonic Flow 93 9 Block BASIC FUNCTION Functional description ➞ ➞ BASIC FUNCTION SENSOR DATA (CH1...CH2) SENSOR PARAMETER SENSOR Use this function to select the configuration for the ultrasonic sensors, e.g. the number of CONFIGURATION traverses (in the clamp-on design) or whether single-path or dual-path configuration is (6882) present (in the insertion design).
Page 122
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ BASIC FUNCTION SENSOR DATA (CH1...CH2) SENSOR PARAMETER WIRE LENGTH The wire length for assembling the sensors at the correct distance apart appears on the (6885) display.
Device functions Proline Prosonic Flow 93 9 Block BASIC FUNCTION 9.4.2 Function group CALIBRATION DATA ⇒ BASIC FUNCTION HART ⇓ PROCESS PARAM. CH1 ⇓ PROCESS PARAM. CH2 ⇓ SYSTEM PARAM. CH1 ⇓ SYSTEM PARAM. CH2 ⇓ ⇒ SENSOR DATA CH1 SENSOR PARAM.
Page 124
9 Block BASIC FUNCTION Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ BASIC FUNCTION SENSOR DATA (CH1...CH2) CALIBRATION DATA DEVIATION Use this function to enter a deviation value for the arc length. ARC LENGTH (6895) Note! This function is not available unless INSERTION was set in the function MEASUREMENT (6880) and the DUAL PATH option was selected in the function SENSOR CONFIGURATION (6882).
Note! For further information on the functional concept of the “Advanced Diagnostics” menu please refer to the Operating Instructions Proline Prosonic Flow 93, BA 070D/06/en/, chapter “Commissioning” REFERENCE Use this function to start determining the user reference status. The following values are...
Page 127
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ CONFIGURATION WARNING MODE Use this function to determine whether a warning should be generated when there is a (7603) deviation between the reference status (USER, see function SELECT REFERENCE CON- DITION (7602)) and the current measuring values.
10 Block SPECIAL FUNCTION Device functions Proline Prosonic Flow 93 10.1.2 Function group ACQUISITION ⇒ SPECIAL FUNCTION H ADVANCED DIAGNOSTICS CH1 ⇓ ⇒ CONFIGURATION ⇓ ADVANCED DIAGNOSTICS CH2 ⇓ ACQUISITION ADVANCED DIAGNOSTICS AVG Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ ACQUISITION...
Page 129
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ ACQUISITION RESET HISTORY Use this function to delete all history values. (7613) Option: NO - YES Factory setting: Endress+Hauser...
10 Block SPECIAL FUNCTION Device functions Proline Prosonic Flow 93 10.1.3 Function group VOLUME FLOW ⇒ SPECIAL FUNCTION H ADVANCED DIAGNOSTICS CH1 ⇓ ⇒ CONFIGURATION ⇓ ADVANCED DIAGNOSTICS CH2 ⇓ ACQUISITION ⇓ ADVANCED DIAGNOSTICS AVG VOLUME FLOW Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ VOLUME FLOW The volume flow can be monitored in this function group and compared with a reference value.
Page 131
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ VOLUME FLOW WARNING LEVEL Note! (7636) This function is not available unless ON was selected in the WARNING MODE (7603) function.
10 Block SPECIAL FUNCTION Device functions Proline Prosonic Flow 93 10.1.4 Function group FLOW VELOCITY ⇒ SPECIAL FUNCTION H ADVANCED DIAGNOSTICS CH1 ⇓ ⇒ CONFIGURATION ⇓ ADVANCED DIAGNOSTICS CH2 ⇓ ACQUISITION ⇓ ADVANCED DIAGNOSTICS AVG VOLUME FLOW ⇓ FLOW VELOCITY Functional description SPECIAL FUNCTION ➞...
Page 133
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ FLOW VELOCITY WARNING LEVEL Note! (7646) This function is not available unless ON was selected in the WARNING MODE (7603) function.
10 Block SPECIAL FUNCTION Device functions Proline Prosonic Flow 93 10.1.5 Function group SIGNAL STRENGTH ⇒ SPECIAL FUNCTION H ADVANCED DIAGNOSTICS CH1 ⇓ ⇒ CONFIGURATION ⇓ ADVANCED DIAGNOSTICS CH2 ACQUISITION ⇓ VOLUME FLOW ⇓ FLOW VELOCITY ⇓ SIGNAL STRENGTH Functional description SPECIAL FUNCTION ➞...
Page 135
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ SIGNAL STRENGTH WARNING LEVEL Note! (7656) This function is not available unless ON was selected in the WARNING MODE (7603) function.
10 Block SPECIAL FUNCTION Device functions Proline Prosonic Flow 93 10.1.6 Function group SOUND VELOCITY ⇒ SPECIAL FUNCTION H ADVANCED DIAGNOSTICS CH1 ⇓ ⇒ CONFIGURATION ⇓ ADVANCED DIAGNOSTICS CH2 ⇓ ACQUISITION ⇓ ADVANCED DIAGNOSTICS AVG VOLUME FLOW ⇓ FLOW VELOCITY ⇓...
Page 137
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS (CH1,CH2,AVG) ➞ SOUND VELOCITY ACTUAL DEVIATION This function displays deviation between the measured sound velocity and the reference (7665) values (USER), see Page 126, selected in the function SELECT REFERENCE CONDITION (7602).
10 Block SPECIAL FUNCTION Device functions Proline Prosonic Flow 93 10.1.7 Function group ACTUAL TRANSIT TIME ⇒ SPECIAL FUNCTION H ADVANCED DIAGNOSTICS CH1 ⇓ ⇒ CONFIGURATION ⇓ ADVANCED DIAGNOSTICS CH2 ACQUISITION ⇓ VOLUME FLOW ⇓ FLOW VELOCITY ⇓ SIGNAL STRENGTH ⇓...
Page 139
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS CH1 ➞ ACTUAL TRANSIT TIME WARNING LEVEL Note! (7676) This function is not available unless ON was selected in the WARNING MODE (7603) function.
10 Block SPECIAL FUNCTION Device functions Proline Prosonic Flow 93 10.1.8 Function group ACCEPTANCE RATE ⇒ SPECIAL FUNCTION H ADVANCED DIAGNOSTICS CH1 ⇓ ⇒ CONFIGURATION ⇓ ADVANCED DIAGNOSTICS CH2 ACQUISITION ⇓ VOLUME FLOW ⇓ FLOW VELOCITY ⇓ SIGNAL STRENGTH ⇓...
Page 141
Device functions Proline Prosonic Flow 93 10 Block SPECIAL FUNCTION Functional description SPECIAL FUNCTION ➞ ADVANCED DIAGNOSTICS CH1 ➞ ACCEPTANCE RATE ACTUAL DEVIATION This function displays deviation between the measured acceptance rate and the reference (7685) values (USER), see Page 126, selected in the function SELECT REFERENCE CONDI- TION (7602).
Device functions Proline Prosonic Flow 93 11 Block SUPERVISION 11.1 Group SYSTEM (SYSTEM CH2) 11.1.1 Function group CONFIGURATION ⇒ ⇒ SUPERVISION SYSTEM CONFIGURATION Functional description ➞ ➞ SUPERVISION SYSTEM CONFIGURATION ASSIGN SYSTEM ERROR Use this function to view all system errors and the associated error categories (8000) (fault message or notice message).
Page 144
11 Block SUPERVISION Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ SUPERVISION SYSTEM CONFIGURATION ERROR CATEGORY Use this function to define whether a process error triggers a notice message or a fault (8003) message. If you select “FAULT MESSAGES”, all outputs respond to an error in accord- ance with their defined error response patterns.
Device functions Proline Prosonic Flow 93 11 Block SUPERVISION 11.1.2 Function group OPERATION ⇒ ⇒ SUPERVISION SYSTEM CONFIGURATION ⇓ ⇓ OPERATION ⇒ SYSTEM CH2 OPERATION Functional description ➞ ➞ SUPERVISION [CH2] SYSTEM OPERATION ACTUAL SYSTEM Use this function to check the present system condition.
Page 146
11 Block SUPERVISION Device functions Proline Prosonic Flow 93 Functional description ➞ ➞ SUPERVISION [CH2] SYSTEM OPERATION VALUE SIMULATION Note! MEASURAND This function is available in groups SYSTEM+SYSTEM CH2. (8044) Note! The function is not visible unless the function SIMULATION MEASURAND (8043) is active.
Device functions Proline Prosonic Flow 93 11 Block SUPERVISION 11.2 Group VERSION INFO 11.2.1 Function group DEVICE ⇒ SUPERVISION SYSTEM ⇓ ⇒ VERSION INFO DEVICE Functional description ➞ ➞ SUPERVISION VERSION INFO DEVICE DEVICE SOFTWARE Displays the current device software version.
• The language options of the available language group are displayed in the LANGUAGE (2000) function. • You can change the language group via the configuration software ToF Tool - Fieldtool Package. Please do not hesitate to contact your Endress+Hauser sales office if you have any questions. Endress+Hauser...
Device functions Proline Prosonic Flow 93 11 Block SUPERVISION 11.2.4 Function group F-CHIP ⇒ SUPERVISION SYSTEM ⇓ ⇒ VERSION INFO DEVICE ⇓ SENSOR ⇓ AMPLIFIER ⇓ F-CHIP Functional description ➞ ➞ SUPERVISION VERSION INFO F-CHIP STATUS F-CHIP Use this function to check whether an F-CHIP is installed.
Device functions Proline Prosonic Flow 93 14 Index key words Index key words Additional line (Multiplex) ..... . 39 Current output....... 54 Frequency (Pulse/freq.
Page 158
14 Index key words Device functions Proline Prosonic Flow 93 Totalizer (1...3)....... 49 Channel 2 .
Page 159
Device functions Proline Prosonic Flow 93 14 Index key words Operation Current output....... 64 Pulse/frequency output .
Page 160
14 Index key words Device functions Proline Prosonic Flow 93 Additional line ....... 39 Information line .
Page 161
Device functions Proline Prosonic Flow 93 14 Index key words Liquid ........117 Operation.
Page 162
14 Index key words Device functions Proline Prosonic Flow 93 Zero point adjustment ......111 Numerics 0_4 mA value .