DISPlay Subsystem Commands

DISPlay[:WINDow[1]]:ANNotation[:ALL] ON|OFF|0|1
DISPlay[:WINDow[1]]:ANNotation[:ALL]?
Turns the screen annotation
labeling and labeling within the graticule.
DISPlay[:WINDow[1]]:TEXT:CLEar
Erases all text on the display resulting from previous use of the DISPlay[:WIN-
Dow[1]]:TEXT:DATA command.
DISPlay[:WINDow[1]]:TEXT:DATA <string>|<data_block>
DISPlay[:WINDow[1]]:TEXT:DATA?
Writes text on the display in the Title area. Use the <data_block> parameter to
send extended ASCII characters such as control codes and symbols.
DISPlay[:WINDow[1]]:TRACe:ALL[:SCALe][:AUTO]
Finds the largest input signal using trace A and sets the span and vertical scale
to display that signal. This command performs the same function as the front-
panel Auto-Meas key.
The following defines the instrument state settings altered by Auto Measure.
State settings that are not listed are not altered.
Because many instrument state setting are altered, it is recommended you use
this command only to find unknown signals. It is not recommend this com-
mand be used in the middle of a measurement routine.
Center Wavelength
Span
Grating Order
or
. Affects only the X-axis and Y-axis
ON
OFF
According to signal wavelength and bandwidth
Set according to automeasure setup panel. In some cases, this
may also be a function of signal characteristics.
Auto
Remote Operation