The Format Menu

Setup/Hold Field (State only)

Setup/Hold Field (State only)
Setup/Hold adjusts the relative position of the clock edge with respect to the
time period that data is valid. When the Setup/Hold field is selected, a
configuration menu appears. Use this Setup/Hold configuration menu to
select each pod in the analyzer and assign a Setup/Hold selection from the
selection list.
With a single clock edge assigned, the choices range from 3.5 ns Setup/0.0 ns
Hold, to 0.0 ns Setup/3.5 ns Hold. With both edges of a single clock assigned,
the choices are from 4.0 ns Setup/0.0 ns Hold, to 0.0 ns Setup/4.0 ns Hold. If
the analyzer has multiple clock edges assigned, the choices range from 4.5 ns
Setup/0.0 ns Hold, to 0.0 ns Setup/4.5 ns Hold.
Setup and hold menu
Setup / Hold field
Setup and Hold Menu
4–22