Access P388 and enter EDIT mode
P388-01 Initial Span
Enter 1 to start Initial Span
Press the following keys:
Span Calibration. Current Span
Setup test. Press ENTER to start
For example: the current span count.
Zero should be done prior to Span
Setup test. Press ENTER to start
Perform a zero calibration or clear.
Press the following key:
Initial Span Calibration in progress
Current Reading:
The span count being calculated while calibration is in progress.
Calibration complete. Deviation
Press ENTER to accept value
The deviation is reset. For example, the new span value if accepted or unacceptable, press the follow-
ing key to restart:
Press the following key:
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
Recalibration
7.4 Recalibration
E
0
41440
0
#####
0.00
41900
79