You cannot override the Safety Pretest warning because the
insulation Safety Pretest is active. If the insulation Safety Pretest
detects a connected appliance, the test sequence stops.
9.
When the test is done, reset the RCD.
10. Use
to review test results. The first result shown is the last measurement
taken, the insulation N-PE test. Press the down arrow
to the first test, the Line test.
Test results are in temporary memory. If you want to store the test results,
press . For more information, see Memory Mode.

Memory Mode

You can store up to 3000 measurements on the Tester. The information stored
for each measurement consists of the test function and all user selectable test
conditions.
The location identifier includes a location set number (a), location subset
number (b), and location ID number (c). You can store multiple measurements
to the same memory location (a, b, c) and view later with the Tester or a
software program such as Fluke DMS Software. With DMS you have additional
tools to apply custom labels to these memory locations. See the DMS Software
User Manual for more information.
h
To enter Memory mode:
1.
Press  to enter Memory mode. See Figure 15.
Use the location set field (a) to indicate a location such as a room or
electrical panel number.
Use the location subset field (b) for circuit number.
Use the location ID field (c) for a socket or place number.
Electrical Installation Tester
Note
Memory Mode
to move backward
63