Maintenance
To prevent possible electrical shock, fire, or personal injury:
•
Be sure that the battery polarity is correct to prevent battery
leakage.
•
Repair the Product before use if the battery leaks.
•
Have an approved technician repair the Product.
•
Use only specified replacement parts.
•
Replace a blown fuse with exact replacement only for
continued protection against arc flash.
•
Do not operate the Product with covers removed or the case
open. Hazardous voltage exposure is possible.
•
Remove the input signals before you clean the Product.
Periodically wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
1.
Turn the meter off and remove all test leads.
2.
Shake out any dirt that may be in the terminals.
3.
Moisten a clean cotton swab with alcohol and clean the inside of each
terminal.
Table 20 is a list of replaceable parts for the Tester.
W Fuse, 11 A, 1000 V 10.3 x 25.4 mm for Fused Probe
W Fuse, 3.15 A, 500 V 6.35 x 32 mm for 166X Tester
XW Warnings
Table 20. Replacement Parts
Description
Electrical Installation Tester
Maintenance
Part Number
803293
2030852
69