4.3.1 Scaling with Known Loads (On-Line Calibration)
For maximum resolution, find the maximum signal that will be applied to the
Note
meter input.
• For regular voltage input, refer to the main body of Table 4-1.
• For millivolt or milliamp input, refer to the main body of Table 4-2.
Set the DIP switch positions as indicated at the top of either Table 4-1 or 4-2. The
numbers 1 through 8 in the top row of either table represent dip switches 1 through 8,
and the O, C or X directly below the number indicates the correct position of each switch.
'O' Switch should be open or up.
'C' Switch should be closed or down.
'X' Switch is used to control excitation (refer to Table 2-3 to determine
correct position of these switches).
Once Dip switches have been positioned correctly, apply power. Proceed to the RD.CF
(Reading Configuration) and set R2 equal to the value in the right hand column of the
chart.
Table 4-1. Range Selection Dip Switch Positions For Regular Voltage Input
12345678
XCOOOXXC
Table 4-2. Range Selection Dip Switch Positions For Millivolt/ Milliamp Input
12345678
Table 4-1. Range Selection Dip Switch Positions For Regular Voltage Input
XOCOOXX0
0 - 100 mV
0 - 50 mV
0 - 30 mV
Table 4-2. Range Selection Dip Switch Positions
0 - 20 mV
0 - 10 mV
p
* Reading Configuration
12345678
XCOOCXXC
0 - 10 V
0 - 5 V
0 - 3 V
0 - 2 V
0 - 1 V
12345678
XOCOCXX0
±50 mV
±50 mV
±30 mV
±20 mV
For Millivolt and Milliamp Input
±10 mV
Configuring The Meter
RD.CF*
±5 V
±5 V
±3 V
±2 V
±1 V
12345678
XOCCOXX0
0 - 20 mA
0 - 10 mA
0 - 6 mA
0 - 4 mA
0 - 2 mA
R.2=
4
3
2
1
0
RD.CF*
R.2=
4
3
2
1
0
4
23