Connecting Sensor Inputs; Meter-Powered Bridge Input; Externally-Powered Bridge Input - Omega DP25B-S User Manual

Strain gage panel meter
Hide thumbs Also See for DP25B-S:
Table of Contents

Advertisement

3.4 CONNECTING SENSOR INPUTS

Figures 3-6 shows excitation supplied from the meter's internal supply
(50 mA maximum) Select 5, 10, or 12 volt excitation at DIP switch.
Figures 3-7 shows the connections required for an externally-powered bridge input: the
external supply is brought to the meter's buffer circuits to permit ratiometric readings.
Set S1 DIP switch for external excitation for Figure 3-7 and 3-8.
+
EX
-
EX
Figure 3-7. Externally-powered Bridge Input
-
+
+
+
E
E
20mA
R
1
2
3
4
TB2
+
+
-
-
24
S
S
R
5
6
7
8
+
-
Figure 3-6. Meter-powered Bridge Input
-
+
+
+
E
E
20mA
R
1
2
3
4
TB2
+
+
-
-
24
S
S
R
5
6
7
8
+
-
Getting Started
Connections with "typical wire colors"
+E = Positive Excitation (red)
-E = Negative Excitation (black)
+S = Positive Signal Input (green)
-S = Negative Signal Input (white)
Connections with "typical wire colors"
+E = Positive Excitation (red)
-E = Negative Excitation (black)
+S = Positive Signal Input (green)
-S = Negative Signal Input (white)