8.2.4 COD/BOD/LPI/FLOW Program
DISPLAY
COD/BOD PROGRAM
STREAM 1
TOC FACTOR
STREAM 2
TOC FACTOR
STREAM 3
TOC FACTOR
........................
STREAM 6
TOC FACTOR
LPI PROGRAM
STREAM 1
LPI VALUE
STREAM 2
TOC FACTOR
STREAM 3
TOC FACTOR
........................
STREAM 6
TOC FACTOR
NO
DISPLAY function, which is programmed as "NO" by default,
allows the user to program COD, BOD, LPI and/or FLOW & e.g.
TOC in kg/h parameter results to be displayed by the system
with the "YES" setting. When activated, the system sends the
relevant 4-20mA signals for COD, BOD (in mgO/l), LPI (in %),
and/or FLOW (in m
output channels if programmed. See Section
Program
In Stream 1 1.0 0.0 setting, the first setting "1.0" acts as the
"OVERALL FACTOR" and the second setting acts as the
1.0
0.0
"OFFSET FACTOR" in the following equation:
1.0
COD (and/or BOD) = [Overall Factor * (TOC Factor * TOC)] +
1.0
0.0
Offset Factor
1.0
In TC analysis mode, instead of the TOC FACTOR, the TC
1.0
0.0
FACTOR is displayed in this menu and used in above equation.
1.0
.........
......
If the DISPLAY is programmed as COD and/or BOD above, the
1.0
0.0
system multiplies the TOC result of the corresponding stream
1.0
(STREAM 1-6) with the TOC FACTOR and with the OVERALL
FACTOR and adds the OFFSET FACTOR to calculate the COD
and/or BOD value.
The STREAM FACTOR for each stream is obtained following
the procedures described in information sheet "I030. TOC to
COD or BOD Correlation Method", which is available in the
MMC/SD card shipped with the BioTector.
Stream 1 factor is used for manual samples and calibration
standards.
LPI Value (Lost Product Index Value) is the application
dependent TOC value for the target lost product. For example
the LPI value for milk in a dairy plant can be determined and
0.0mgC/l
installed as 60000 mgC/l.
When the display of the LPI values is activated with the
0.0mgC/l
DISPLAY function above, BioTector calculates the % LPI values
using the measured TOC result and the site specific LPI value
0.0mgC/l
.........
......
programmed in this menu as follows:
LPI (%) = [ (TOC Result) / (LPI VALUE) * 100]
0.0mgC/l
When activated, the % LPI values are displayed in analysis
result screen and in the reaction archive menus.
If the DISPLAY of the Flow Program is activated in systems built
with the sample flow analog input signal option, in addition to the
% LPI value, BioTector calculates the LP Value (Lost Product in
l/h) using the measured TOC result, the specific LPI Value and
the recorded stream specific Sample Flow value as follows:
LP (l/h) = [ (TOC Result) / (LPI VALUE) * Sample Flow * 1000]
When activated, the LP values in l/h are displayed in analysis
result screen and in the reaction archive menus.
/h) & e.g. TOC (in kg/h) to the relevant result
3
for details.
8.2.8
4-20mA