Using The Flow Meter - Hach 9186 Instrument Manual

Oxygen scavenger analyzer
Table of Contents

Advertisement

Installation
Figure 17
Terminal Strip And Circuit Board Descriptions
3
2
1
1.
K5- Not used with Hydrazine
2.
J1- Working Cell and Counter Electrode Connections
3.
Analog Outputs
4.
P1 (contrast adjustment)

4.6 Using the Flow Meter

4.7 Powering On the Analyzer
22
4
K2
P1
Temp +
Temp –
Ref
Counter
Work
Aux
RP2
J1
K6
K5
9180
AMP
V1.03
Ic4
RP1
K4
IC2
When all plumbing and electrical connections are made, slowly open the supply
valve to the inlet tube. Rotate the dial marked M to adjust the flow rate through the
analyzer. The flow rate MUST be stable, and between 10-15 L/hr. Fluctuations in
the flow rate will cause fluctuations in the reading on the display.
Confirm that the site voltage corresponds to that stated on the instrument
identification plate. Turn on power to the transmitter. The transmitter performs a
self-test and displays an initial value.
5
J1
RE3
IC1
C8
L2
Model : 09125-A-2000
Serial No : 0104 549
C6
K9
Vac : 100 - 240 V
Z6
Wiring located behind the guide plate shall be
rated at 80 C (176 F) or higher.
5.
K9-Internal Shield Connections
6.
Fuse Holder (F1, T, 2A, L, 250V)
7.
S1-S4 (relays, alarm and timer connections)
8.
ac Power Connections
6
RE1
V1
V2
V3
V4
Ind : D
7
8

Advertisement

Table of Contents
loading