Sensor Mag 1100 Food - Siemens MAG 1100 Handbook

Electromagnetic flowmeters
Table of Contents

Advertisement

SITRANS F M MAGFLO

2.2 Sensor MAG 1100 FOOD

Type
mm
Nominal size
Process connection
Operating pressure
Vacuum
Temperature of medium
Temperature shock
Ambient temperature
Liner
Electrodes
Enclosure
Terminal box
Standard
(not compact)
Option
Cable entries
Standard
Enclosure rating
Option
Mechanical load (vibration)
Test pressure
Approvals
Excitation frequency
Conforms to PED, LVT, EMC
Accessories
MAG 1100 FOOD
2. Technical data
MAG 1100 FOOD
♦ Direct welding into dairy pipe ♦ Clamp fitting ♦ Threaded fitting
DN 10-65: 40 bar, DN 80: 37.5 bar, DN 100: 30 bar
1 × 10
-6
bar
−20°C to +150°C
Suitable for steam sterilization
(Duration > 1 min.):
Max. ∆T ≤ 15°C/min.
DN 10, 15, 25
Max. ∆T ≤ 10°C/min.
DN 40, 50, 65
Max. ∆T ≤ 5°C/min.
DN 80, 100
(Duration ≤ 1 min., followed by 10 min. rest):
Max. ∆T ≤ 80°C
DN 10, 15, 25
Max. ∆T ≤ 70°C
DN 40, 50, 65
Max. ∆T ≤ 60°C
DN 80, 100
−40°C to +100°C
Remote signal converter:
Compact signal converter: −20°C to +50°C
Aluminium oxide Al
O
(ceramic)
2
3
Platinum with gold/titanium brazing alloy
Stainless steel AISI 316 L (1.4404)
Fibre glass-reinforced polyamide
Stainless steel AISI 316 (1.4436)
4 Pg 13.5
IP 67 to EN 60529 (NEMA 4x) (1 m w.g for 30 min.)
IP 68 to EN 60529 (NEMA 6) (10 m w.g. cont.)
18-1000 Hz random, 3.17 G rms in all directions
to EN 60068-2-36
80 bar (2 × PN)
3A, EHEDG
DN 10-65: 12.5 Hz
DN 80-100: 6.25 Hz
PED - 97/23EC, LVD - 73/23 EEC + amendment 93/68/EEC, EMC - 89/336 EEX
Adapter
Stainless steel AISI 316
Pipe connection/
Adapter for direct welding into dairy pipe:
Tri-Clover ISO 2037, DIN 11850, SMS 3008, BS 4825-1
Operating
pressure
Clamp adapter:
Tri-Clamp ISO 2852, DIN 32676, SMS 3016, BS 4825-3
Thread adapter:
DIN 11851:
ISO 2853, SS 3351, BS 4825-4:
SMS 1145:
EPDM (−20 °C to 150 °C)
Gasket Standard
Option NBR (−20 °C to 100 °C)
Clamp
Stainless steel AISI 304, ISO 2852
Note
When combined sensor and adapter, the working pressure is the lower rated of the pair.
DKFD.PS.027.W3.02
Hygienic sensor
DN 10, 15, 25, 40, 50, 65, 80, 100
Hygienic adapters available for:
20 bar
0.02 bar
−30°C to +130°C
Suitable for steam sterilization at 150°C
Max. ±100°C momentarily
Remote signal converter:
Compact signal converter: −20°C to +50°C
Reinforced PFA (Teflon)
Hastelloy C-276
Stainless steel AISI 316 L (1.4404)
Fibre glass-reinforced polyamide
Stainless steel AISI 316 (1.4436)
4 Pg 13.5
IP 67 to EN 60529 (NEMA 4x) (1 m w.g for 30 min.)
IP 68 to EN 60529 (NEMA 6) (10 m w.g. cont.)
18-1000 Hz random, 3.17 G rms in all directions
to EN 60068-2-36
40 bar (2 × PN)
3A
DN 10-65: 12.5 Hz
DN 80-100: 6.25 Hz
DN 10, 15, 25, 40, 50, 65, 80
DN 100
DN 10, 15, 25, 40, 50
DN 65, 80, 100
DN 10, 15, 25, 40
DN 50, 65, 80, 100
DN 10, 15, 25, 40, 50, 65, 80
DN 25, 40, 50, 65, 80
MAG 1100 FOOD
MAG 1100 FOOD PFA
−40°C to +100°C
Pressure
PN 40
PN 25
PN 16
PN 10
PN 40
PN 25
PN 16
PN 6
9