7. Electrical connection

7.1
Signal converter
MAG 5000 and MAG 6000
connection diagram
Note
Special cable with individual
wire shields (shown as
dotted lines) are only
requried when using empty
pipe function or long cables.
(See "Technical data"
chapter 2 for further details.)
7. Electrical connection
Grounding
PE must be connected due to safety class 1 power supply.
Mechanical counters
When mounting a mechanical counter to terminals 57 and 58 (active output), a 1000 µF capacitor
must be connected to the terminals 56 and 58.
Capacitor + is connected to terminal 56 and capacitor − to terminal 58.
Output cables
If long cables in noise environment, we recommend to use screened cable.
Electrodes cables
Dotted connections only to be when using special electrode cable.
DKFD.PS.027.W3.02