Installation Of Sensor; Potential Equalization - Siemens MAG 1100 Handbook

Electromagnetic flowmeters
Table of Contents

Advertisement

SITRANS F M MAGFLO

5. Installation of sensor

5.1

Potential equalization

MAG 1100
MAG 1100 FOOD
MAG 3100 W / MAG 3100
(except PTFE liner)
5. Installation of sensor
To obtain optimum results from the measuring system, the chassis body of the sensor must have
the same electrical potential as the liquid being measured.
Graphite gaskets
Electrically
conductive
piping
A: Potential equalization with electri-
cally conductive graphite gaskets
Electrically
non-conductive
piping
C: Potential equalization with electri-
cally conductive graphite gaskets
DKFD.PS.027.W3.02
EPDM or PTFE gaskets
B: Potential equalization using earth
strap supplied.
D: Potential equalization using sepa-
rate potential equalization ring
The sensor must be installed between two adap-
ters. Potential equalization with the liquid oc-
curs automatically via these adapters and
through the adjacent pipe.
Potential equalization is carried out with the
built-in earthing electrodes. No further action
need to be taken.
37