Maintenance; Performance Specifications; Environment Specifications - A&D UA-767PC Instruction Manual

Hide thumbs Also See for UA-767PC:
Table of Contents

Advertisement

Do not open the case of the device because it uses delicate electrical
components and an intricate air unit that could be damaged. If you can not locate
and fix the problem, request service from your supplier, or from the A&D service
group. The A&D service group will support authorized suppliers about technical
information, spare parts and units.
The technical testing procedures – to be carried out at least every 3 years – can
be performed either by the manufacturer or by an authorized repair service in
accordance with the regulations governing manufactures of medical products.

Performance Specifications

Measurement method Oscillometric measurement
Measurement range
Accuracy
Battery
Cuff
Classification
Clinical test

Environment Specifications

Operating environment +10 °C - +40 °C, less than 85 %RH
Storage environment
Physical Specifications
Dimensions
Weight

Maintenance

Specifications
Pressure : 20 - 280 mmHg
Pulse :
40 - 200 beats / minute
Pressure :±3 mmHg or 2%, whichever is greater
Pulse :
±5%
4 x 1.5V (R6P, LR6, or AA)
Arm Circumference
Type BF
Device, cuff and tubing are designed to provide special
protection against electric shocks.
According to ANSI / AAMI SP-10 1987
Systolic blood pressure
Diastolic blood pressure
−10 °C - +60 °C, less than 85 %RH
163.7 [W] x 111 [D] x 66.7 [H] mm
Approx. 320g without batteries
0366
Standard deviation
5.43 mmHg
5.18 mmHg
15
Mean error
2.24 mmHg
1.54 mmHg