Analog Input
AI <0..n>
Mux
AI SENSE
AI GND
AI Ground-Reference
© National Instruments Corporation
Figure 4-1 shows the analog input circuitry of USB-621x devices.
DIFF, RSE,
NI-PGIA
or NRSE
Input Range
Selection
Settings
Figure 4-1. USB-621x Analog Input Circuitry
The main blocks featured in the USB-621x analog input circuitry are as
follows:
•
I/O Connector—You can connect analog input signals to the
USB-621x device through the I/O connector. The proper way
to connect analog input signals depends on the analog input
ground-reference settings, described in the
Ground-Reference Settings
Device-Specific
•
Mux—Each USB-621x device has one analog-to-digital converter
(ADC). The multiplexers (mux) route one AI channel at a time to the
ADC through the NI-PGIA.
•
AI Ground-Reference Settings—The analog input ground-reference
settings circuitry selects between differential (DIFF), referenced
single-ended (RSE), and non-referenced single-ended (NRSE) input
modes. Each AI channel can use a different mode.
•
NI-PGIA—The NI programmable gain instrumentation amplifier
(NI-PGIA) is a measurement and instrument class amplifier that
ADC
AI FIFO
section. Also refer to Appendix A,
Information, for device I/O connector pinouts.
4-1
4
Isolation
Barrier
(USB-6215/
6216/6218
devices only)
Digital
AI Data
Isolators
Analog Input
NI USB-621x User Manual