Analog Input

The NI PXIe-7862 provides connections for 16 AI channels. Each channel has an AI+ pin, AI-
pin, and AIGND pin to which you can connect both single-ended or differential voltage
signals. Use the AISENSE pin to connect non-referenced single-ended signals.

Connecting Single-Ended Voltage Signals

To connect referenced single-ended voltage signals to the NI PXIe-7862, you must connect the
voltage ground signal to AI GND in order to keep the common-mode voltage in the specified
range.
Figure 2. Connecting Referenced Single-Ended Signals to the NI PXIe-7862
+
V
1
To connect non-referenced single-ended voltage signals to the NI PXIe-7862, you must
connect the voltage ground signal to AI SENSE in order to keep the common-mode voltage in
the specified range.
Figure 3. Connecting Non-Referenced Single-Ended Signals to the NI PXIe-7862
+
V
1
+
V
cm
4 | ni.com | NI PXIe-7862 User Manual
AI+
AI–
AISENSE
AIGND
Connection
Accessory
AI+
AI–
AISENSE
AIGND
Connection
Accessory