Chapter 4
Analog Input
AI Ground-Reference
*
Setting
Differential (DIFF)
Non-Referenced
Single-Ended (NRSE)
Referenced Single-Ended
(RSE)
*
Refer to the
Analog Input Ground-Reference Settings
software considerations.
Refer to the
Connecting Ground-Referenced Signal Sources
NI USB-621x User Manual
Table 4-2. USB-6210/6211/6212 Analog Input Configuration

Floating Signal Sources

(Not Connected to
Building Ground)
Examples:
• Ungrounded thermocouples
• Signal conditioning with
isolated outputs
• Battery devices
Signal Source
+
Signal Source
+
Signal Source
+
Example:
• Plug-in instruments with
DAQ Device
AI+
+
AI–
AI GND
DAQ Device
AI
+
AI SENSE
AI GND
DAQ Device
AI
+
AI GND
section for descriptions of the RSE, NRSE, and DIFF modes and
section for more information.
4-24
Ground-Referenced
Signal Sources
non-isolated outputs
DAQ Device
Signal Source
AI+
+
AI–
AI GND
V cm
Signal Source
DAQ Device
AI
+
AI SENSE
AI GND
V cm
NOT RECOMMENDED for the
USB-6210/6211/6212
Signal Source
DAQ Device
AI
+
V
A
AI GND
V
B
Ground-loop potential (V
– V
) are added
A
B
to measured signal.
+
+
+