With this type of connection, the NI-PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the signal
source and the device ground, shown as V
AI+ and AI– must both remain within ±11 V of AI GND.
Using Non-Referenced Single-Ended (NRSE)
Connections for Ground-Referenced Signal Sources
Figure 4-23 shows how to connect ground-reference signal sources to the
USB-6210/6211/6212 device in NRSE mode.
AI +
AI –
Input Multiplexers
AI SENSE
AI GND
USB-6210/6211/6212 Configured in DIFF Mode
cm
4-34
Instrumentation
Amplifier
+
+
PGIA
Measured
V
m
Voltage
in the figure.