Analog Input Timing Signals

20 MHz Timebase
100 kHz Timebase
© National Instruments Corporation
In order to provide all of the timing functionality described throughout this
section, USB-621x devices have a flexible timing engine. Figure 4-4
summarizes all of the timing options provided by the analog input timing
engine.
AI Sample
Clock
Timebase
AI Convert
Clock
Timebase
Figure 4-4. Analog Input Timing Options
USB-621x devices use AI Sample Clock (ai/SampleClock) and AI Convert
Clock (ai/ConvertClock) to perform interval sampling. As Figure 4-5
shows, AI Sample Clock controls the sample period, which is determined
by the following equation:
1/Sample Period = Sample Rate
Analog Comparison Event
Ctr n Internal Output
SW Pulse
Programmable
Clock
Divider
Ctr n Internal Output
Programmable
Clock
Divider
4-11
Chapter 4
Analog Input
AI Sample Clock
AI Convert Clock
NI USB-621x User Manual