Analog Input

Figure 4-1 shows the analog input circuitry of NI 6238 and NI 6239
devices.
AI <0..7>+
Mux
AI Terminal
Configuration
AI <0..7>–
Selection
AI GND

Analog Input Circuitry

I/O Connector
You can connect analog input signals to the M Series device through the I/O
connector. The proper way to connect analog input signals is outlined in the
Connecting Analog Current Input Signals
Appendix A,
pinouts.
MUX
Each M Series device has one analog-to-digital converter (ADC). The
multiplexers (MUX) route one AI channel at a time to the ADC through the
NI-PGIA.
Instrumentation Amplifier (NI-PGIA)
The NI programmable gain instrumentation amplifier (PGIA) is a
measurement and instrument class amplifier that minimizes settling times
© National Instruments Corporation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Isolation
NI-PGIA
ADC
Isolators
Input Range
Selection
Figure 4-1. NI 6238/6239 Analog Input Circuitry