14.7 Replacing the batteries

The thermometer is supplied with two 1.5 V batteries type AA (LR 6).
For best performance, we recommend Duracell
1
Insert new batteries when the battery
symbol begins to flash on the display
(See 12. Errors and notifications).
2
Open battery door by sliding spring
latch to the right using your finger or a
pointed object such as a pen. While
holding the latch in the open postion,
grasp the battery door and remove it.
3
Remove the batteries and replace with new batteries, making sure
the poles are in the right direction.
4
Snap battery door back into place and ensure the latch returns to
original latch position.
This product contains batteries and recyclable electronic waste. To protect the
environment, do not dispose of it in the trash, but take it to appropriate local collection
points according to national or local regulations.

14.8 Calibration testing

The thermometer is initially calibrated at the time of manufacture. If the thermometer is used
according to the use instruction, periodic readjustment is not required. However, Welch Allyn
recommends checking calibration on an annual basis or whenever clinical accuracy of the
thermometer is in question. Procedures for checking calibration are outlined in the 9600 Plus
Calibration Tester (REF 01802-110) directions for use manual.
The above recommendations do not supersede the legal requirements. The user must always
comply with legal requirements for the control of the measurement, functionality, and accuracy of
the device which are required by the scope of relevant laws, directives or ordinances where the
device is used.