As needed, clean the thermometer body and cradle following the below directions.
1
For additional protection, we recommend placing a new probe
cover on the thermometer probe to protect this area when
cleaning the body of the thermometer.
2
Use a damp cloth or cleaning wipe with cleaning solution from
the Approved Cleaning Solutions Table. To clean the body,
ensure that the wipe is moist, not saturated. Wipe the body with
the display facing up.
3
Remove probe cover box carrier from the cradle by rotating
the probe cover box carrier forward. See 14.5 Removing and
installing probe cover box carrier
4
Wipe the cradle and probe cover carrier with a damp cloth
or cleaning wipe with cleaning solution from the Approved
Cleaning Solutions Table.
5
Allow at least 5 minutes drying time before taking temperature.
Make sure probe, body and cradle are clean and dry prior to use.