5.1.2.
Service Mode 1 (Function Parameter Setting)
Use the following procedure to change the function parameters.
Step
1
Standby
2
Press "FUNCTION" and then "7".
3
Press "MONITOR" four times, then press "*".
4
Press "1".
5
Enter the Function Parameter Number.
Ex: Changing the "ALARM STATUS" -- Enter "001" and
press [SET].
6
Press "START".
7
Enter the new setting value. Ex: Enter "3" for Constant.
8
Press "START". The new value will be stored and the
next parameter will be displayed.
9
Repeat steps 4 through 7 to change other Function
Parameters or Press "STOP" twice to return to standby.
Note:
The following buttons provide these functions in the Service Mode:
"START"
: The new setting value is stored in the machine.
"V"
: Scroll the function parameter number down.
"/\"
: Scroll the function parameter number up.
Service Mode 1
Operation or Unit Condition
140
LCD Display
MMM-dd-yyyy
15:00
00%
SET MODE
(1-6)
ENTER NO. OR
SERVICE MODE
ENTER NO. OR
PARAMETER (000-199)
ENTER PARAMETER #_
PARAMETER #001
ALARM STATUS?
ALARM STATUS:Timer
1:OFF 2:Tmr 3:CONST
ALARM STATUS:Const.
1:OFF 2:Tmr 3:CONST
PARAMETER #002
STOP COMM.JRNL?
MMM-dd-yyyy
15:00
00%