DISPLAY BACKLIGHT

Press the HOLD/
key for >1 second to turn the backlight on. The backlight will
automatically turn off after 10 seconds.
HOLD
The hold function freezes the reading in the display. Press the HOLD key momentarily to
activate or to exit the HOLD function.
PEAK HOLD
The Peak Hold function captures the peak AC voltage or current. The meter can capture
negative or positive peaks as fast as 1 millisecond in duration. Momentarily press the
PEAK button, "PEAK" and "MAX" will display. The meter will update the dispay each time
a higher positive peak occurs. Press the PEAK button again, "MIN" will display. The meter
will update the dispay each time a lower negative peak occurs. Press and hold the PEAK
button for more than 1 second to exit PEAK Hold mode. Auto Power Off feature will be
disabled automatically in this mode.

LOW BATTERY INDICATION

With a fresh battery installed, the battery icon
the lower right corner of the LCD. The lines will disappear as the battery is used. When the
icon appears alone in the display, the battery should be replaced.

AUTO POWER OFF

The auto off feature will turn the meter off after 15 minutes. To disable the auto power off
feature, hold down the MODE button and turn the meter on. "APO d" will appear in the
display. Turn the meter off and then on again to re-enable the auto power off feature.