3 OPERATING YOUR BALANCE

3.1 The Menu (Basic Settings of the Instrument)

The balance has three basic menus; each is selected by front panel buttons marked Mode, Units and Setup.
Mode Button
The Mode button, when pressed, permits the selection of five weighing modes which are: weigh, percent, count, animal
weighing and checkweigh. These modes are controlled by an on or off selection made in the Setup menu under the
Mode menu.
Units Button
The Units button, when pressed, allows you to select or change units.
Setup Button
The Setup button, when pressed, allows entry into menus which allow you to set the balance for specific operating
parameters. Each of the menus contain settings which are user selectable. The tables below illustrates the various
menus and the functions which are selectable. The items shown on the menu, which are bolded, are the factory default
settings. If the Setup menu is not entered, the balance would function in the basic manner shown by the various
settings which are bolded. The setup menus shown below are arranged in the order as displayed in the balance.
START
CAL (CALIBRATION)
InCal
TM
Calibration (No Lock)*
Span Calibration
User Calibration
Linearity Calibration
Calibration Test (No Lock)
Calibration Now ON/OFF
Calibration Adj ± 100 (0)
Lock ON/OFF
Exit
*(If option is installed)
READOUT
Stable
.5d, 1d, 2d, 5d
Auto 0
OFF, .5d, 1d, 3d
Filter
-0-, -1-, -2-, -3-
Lock
ON/OFF
Exit
PRINT
Auto Print
OFF, Cont, Inter, On Stb
Interval
Enter 1 seconds
Stable
ON/OFF
Numeric
ON/OFF
GLP Con
ON/OFF
GLP Tar
ON/OFF
Reference
ON/OFF
Lock
ON/OFF
Exit
UNITS
Units (grams only)
ON/OFF
Lock
ON/OFF
Exit
SETUP SUBMENUS
DATE
Type m/d/y, dMy, yMd, Myd,ydM,
or dyM
Set
Exit
GLP DATA
User Number
Enter 6 digits
Project Number Enter 6 digits
Lock
ON/OFF
Exit
RS232
Power
ON/OFF (Battery mode only)
Baud 300, 1200, 2400, 4800, 9600
Parity
None, -E-, -Odd-, -0-, -1-
Data 7,8
Stop
1, 2
Lock ON/OFF
Exit
GLOBAL
List
NO/YES
Reset
NO/YES
Version
Software No.
Lock
ON/OFF
Exit
11
TIME
Type 12 Hour/24 Hour
Set Time
Adjust Enter xx seconds
Exit
GLP SET
Time
ON/OFF
Balance ID
ON/OFF
User Number
ON/OFF
Project Number ON/OFF
Difference
ON/OFF
Name
ON/OFF
Lock
ON/OFF
Exit
MODE
Weigh
ON/OFF
Percent
ON/OFF
Count
ON/OFF
Animal
ON/OFF
Checkweigh
ON/OFF
Lock
ON/OFF
Exit
CUSTOM
Factor
Enter 6 digits
Exponent
3 , 2, 1, 0, -1, -2, -3
LSD
.5, 1, 2, 5, 10, 100
Lock
ON/OFF
Exit
AUTO SHUT OFF
Timer
0 - 60 minutes
Lock
ON/OFF
Exit