System Accuracy Results for Glucose
Concentrations <5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±0.28 mmol/L
45.1%
(92/204)
System Accuracy Results for Glucose
Concentrations ≥5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5%
68.9%
(273/396)
Detailed information
about your system
Within
±0.56 mmol/L
85.3%
(174/204)
Within ±10%
94.4%
(374/396)
Within
±0.83 mmol/L
98.5%
(201/204)
Within ±15%
99.2%
(393/396)