Front View; Rear View; Power Connections - Fema S40-T User Manual

Series s40 panel meter 72x36mm size for pt100 / rtd for thermocouples
Hide thumbs Also See for S40-T:
Table of Contents

Advertisement

Front View

Alarms
Logo
Button
Button
LE
UP

Rear View

Option1
Signal
User's Manual S40-T
Button
Units
SQ
Option2
Power
FEMA ELECTRÓNICA - Page 3
Input signal connections - Pt100/RTD
Measure can be selected for 2 or 3 wire systems.
1 2 3 4 5
1 2 3 4 5
Input signal connections - Thermocouple
To configure, select the appropriate thermocouple type in the
configuration menu.
1 2 3 4 5

Power Connections

6 7 8 9 0
Earth connection - Although a terminal is offered for earth connection, the
connection is optional. The instrument does not need this connection for
correct functioning nor for compliance with the security regulations.
Fuse - To comply with security regulation 61010-1, add to the power line
a protection fuse acting as disconnection element, easily accessible to the
operator and identified as a protection device.
Power "H"
fuse 250mA time-lag
Power "L"
fuse 400mA time-lag
Select input Pt100 2Wires in the
menu. Terminal 2 can be left un-
connected.
RTD/Pt100
2 wires
Select input Pt100 3Wires in the
menu
RTD/Pt100
3 wires
+
Thermocouple
-
~
+
-
~

Advertisement