Control Solution Range

If test results fall outside the expect-
ed range, repeat the test. Results that
fall outside the expected range may
indicate the system is not working
properly. Some possible causes are:
Procedural error
I
Expired or contaminated
I
control solution
Test strip code and meter code
I
do not match
Dirty meter or test strip holder
I
Test strip deterioration
I
Meter malfunction
I
I
Control solution that is outside
of the temperature range 15–35˚C
(59–95˚F)
If you are unable to resolve the
problem, call the LifeScan Customer
Care Line at 1 800 663-5521
(U.S.A. 1 800 227-8862).