Quality Assurance / Quality Control Manual
®
Caring for the TRUEresult System
Store the System (Meter, Test Strips, Glucose Control Solution) in an
area protected from liquids, dust and dirt.
Store the System in a dry place at room temperature, 36° - 86°F.
DO NOT REFRIGERATE OR FREEZE.

Meter

Wipe Meter with a clean, lint-free cloth dampened with 10% bleach
and water, or an OSHA approved disinfectant.
Never put Meter in liquids or allow any liquids to enter Test Port.
Do not use 100% Isopropyl alcohol to clean the Meter. It may result
in damage to the Meter Display.

TRUEtest Test Strips

Store Test Strips in original vial only. Do not transfer old Test Strips
into new vial or store Test Strips outside of vial.
Write the date first opened on Test Strip vial label. Discard vial and
unused Test Strips if either EXP date printed on Test Strip vial label
or 4 months after date written on vial label has passed.
Close Test Strip vial immediately after removing one Test Strip. Store
in a dry place at room temperature below 86°F.
DO NOT REFRIGERATE OR FREEZE.
Do not reuse Test Strips.
Do not bend, cut or alter Test Strips in any way.

TRUEtest Glucose Control Solution

Write the date first opened on Control bottle label. Discard bottle if
either EXP date printed on bottle label or 3 months after date written
on bottle label has passed.
After use, wipe bottle tip clean and recap tightly.
Store at room temperature, 36° - 86°F.
DO NOT REFRIGERATE OR FREEZE.