What Display Symbols Mean; Monitor Components - LifeSource UA-704 Mini Manual Inflate

Blood pressure monitor mini manual inflate
Hide thumbs Also See for UA-704:
Table of Contents

Advertisement

?

WHAT DISPLAY SYMBOLS MEAN

Display
Condition/Cause
Symbol
Appears during
measurement; blinks when
MEASURING
pulse is detected.
Appears on display prior to
and during measurement.
Blinks if there is air in cuff
DEFLATE
at power up.
Cuff needs to be inflated for
proper measurement.
INFLATE
An irregular heartbeat or
body movement may have
occurred. Refer to page E-9
IRREGULAR
for more information on
HEARTBEAT
irregular heartbeats.
The monitor is using a fully
charged battery.
FULL BATTERY
The battery is low when it
blinks.
LOW BATTERY
Unstable blood pressure due
to movement during the
measurement.
ERROR
The systolic and diastolic
values are within 10 mmHg
of each other.
The pressure value did not
increase during inflation.
The cuff is not fastened
correctly.
The pulse is not detected
correctly.
PUL. DISPLAY ERROR
LifeSource Health Line (Toll-Free): 1-888-726-9966
?
E-2
Recommended
Action
Remain still and refrain
from talking; measurement
in progress.
Press exhaust button to
release air.
Inflate cuff 30 to 40 mmHg
above your normal systolic
pressure.
Take measurement again
and consult with your
physician.
Replace the battery with
new one, when the
indicator blinks.
Try measurement again.
Remain very still during the
measurement.
Fasten the cuff correctly,
and try the measurement
again.
Try the measurement again.
Remain very still.

MONITOR COMPONENTS

CUFF
SYSTOLIC PRESSURE
DISPLAY
EXHAUST VALVE
INFLATOR BULB
BATTERY COMPARTMENT
BATTERY COVER
In Canada (Toll-Free): 1-800-463-5414
?
AIR TUBE
START BUTTON
DIASTOLIC PRESSURE
OR PULSE DISPLAY
AA BATTERY
?
E-3