What Display Symbols Mean - LifeSource UA-767T Instruction Manual

Talking one step auto-inflation
Hide thumbs Also See for UA-767T:
Table of Contents

Advertisement

~
WHAT DISPLAY SYMBOLS MEAN
Display
Condition/
Recommended
Symbol
Cause
Action
~
Appears during measure-
Remain still and refrain
ment; blinks and beeps
from talking; measure-
MEASUREMENT IN
when pulse is detected.
ment in progress.
PROGRESS
00
Battery voltage is
Replace all batteries
too low for monitor
with new ones.
LOW BATTERY
to work properly.
r
Appears when systolic
Check for air leakage; make
,-
,- ,-
and diastolic measure-
sure cuff tube is properly
-
ments are within 10
connected to monitor
1
mmHg of each other.
and exhaust rate is in
ERROR 1
increments of between
2 mmHg. and 5 mmHg
r
Appears if pressure
Try measurement again
,-
,- ,-
-
2
value is unstable due
and remain very still.
to
movement during
ERROR 2
reading.
r
Appears when pressure
Make sure cuff tube is
,-rr
value did not increase
properly connected to
-
3
during cuff inflation:
monitor, cuff is securely
ERROR 3
1) cuffis not fastened
fastened and exhaust rate
2) air connector is back-
is between 2 mmHg and
wards or not connected
5 mmHg. Make sure air
properly.
connector is inserted
NOTE: If you hear an announcement
properly into tube
indicating that a measurement error
and monitor.
has occurred, take the corrective
MONITOR -
~
-
CUFF HOSE
actions shown above and re-measure.
3
0
LifeSource Health Line (Toll-Free): 1-888-726-9966

Hide quick links:

Advertisement

Table of Contents
loading