Control specifications

1.

Temperature control range

Settable temperature:
Internal temperature display range:
Setting method:
2.

Alarm and safety functions

(1)
Error code
E01: Temp. sensor is open circuited.
When temperature in temp. sensor is equal or higher than -180 , 'E01' is
displayed in 'TopScreen', the basic screen.
E02: Temp. sensor is short circuited.
When temperature in temp. sensor is equal or lower than +50 , 'E02' is
displayed in 'TopScreen', the basic screen.
E03: Cascade sensor is open circuited.
When temperature in cascade sensor is equal or lower than -70 , 'E03' is
displayed in 'TopScreen', the basic screen.
E04: Cascade sensor is short circuited.
When temperature in cascade sensor is equal or higher than +100 , 'E04' is
displayed in 'TopScreen', the basic screen.
E05: Filter sensor is open circuited.
When temperature in filter sensor is equal or lower than -70
displayed in 'TopScreen', the basic screen.
E06: Filter sensor is short circuited.
When temperature in filter sensor is equal or higher than +100 , 'E06' is
displayed in 'TopScreen', the basic screen.
E07: AT sensor is open circuited.
When temperature in AT sensor is equal or lower than -70 , 'E07' is displayed
in 'TopScreen', the basic screen.
E08: AT sensor is short circuited.
When temperature in AT sensor is equal or higher than +100
displayed in 'TopScreen', the basic screen.
E09: When Alarm test is performed with battery uncharged or battery switch is OFF
position, 'E09' is displayed in 'TopScreen', the basic screen.
E10: When temperature in filter sensor is equal or higher than +60
displayed in 'TopScreen', the basic screen, and Compressor H is ceased.
-125
~ -155
-180
In 'TopScreen', press MENU key and step into temperature setting mode
with Shift key (upward, downward, rightward, leftward), then set your
desired value with Figure Input key. Press MENU/OK to store the value.
22
~ +50
, 'E05' is
,'E08' is
,'E10' is