Specifications - Nonin Onyx Vantage 9590 Instructions For Use Manual

Finger pulse oximeter
Hide thumbs Also See for Onyx Vantage 9590:
Table of Contents

Advertisement

Specifications

Oxygen Saturation Display Range:
Pulse Rate Display Range:
Declared Accuracy:
The tables below show A
values measured using the Onyx Vantage 9590 in a clinical study in motion and non-motion conditions.
rms
Decade
70 – 80%
80 – 90%
90 – 100%
70 – 100%
0% to 100% SpO
18 to 321 beats per minute (BPM)
Accuracy Summary by Decade – Finger and Thumb
Oxygen Saturation
Low Perfusion
(A
)
Oxygen Saturation (A
rms
±2
±2
±2
±2
Accuracy Summary by Decade – Toe
Decade
Oxygen Saturation (A
70 – 80%
80 – 90%
90 – 100%
70 – 100%
2
Motion Oxygen Saturation
)
(A
) (Finger)
rms
rms
±2
±3
±2
±3
±2
±2
±2
±3
Low Perfusion
)
rms
Oxygen Saturation (A
±2
±2
±3
±2
±3
±2
±3
±2
This graph shows plots of the error (SpO
by SaO
using the 9590 with a linear regression fit
2
and upper 95% and lower 95% limits of agreement.
Each sample data point is identified by subject
from a clinical study in non-motion conditions.
This graph shows plots of the error (SpO
by SaO
using the 9590 with a linear regression fit
2
and upper 95% and lower 95% limits of agreement.
Each sample data point is identified by subject
from a clinical study in motion conditions.
This graph shows plots of the error (SpO
by SaO
using the 9590 with a linear regression fit
2
and upper 95% and lower 95% limits of agreement.
Each sample data point is identified by subject
from a clinical study using toes in non-motion
conditions.
Motion Oxygen Saturation
(A
) (Thumb)
rms
±3
±4
±3
±3
)
rms
– SaO
)
2
2
– SaO
)
2
2
– SaO
)
2
2
5